Ultrasonic flow meter SONOELIS SE404X,  

SONOELIS SE406X 

 

ELIS PLZE

Ň

 a. s., Lucni 15, P. O. BOX 126, 304

 

26 Plzen, Czech Republic, Phone: +420/377517711, fax: +420/377517722 

Es90201K/c

 

7.3.1.9. Mass + 

See section 7.3.1.6, for the aggregate mass flow in the positive direction. 

7.3.1.10. Mass - 

See section 7.3.1.7, for the aggregate mass flow in the reverse direction. 

7.3.1.11. Temperature 

The temperature reading is only available if a thermometer is included in the meter configuration. The reading 
sensitivity is 0.1 °C. 

7.3.1.12. Density 

The fluid density readings are available provided a thermometer has been installed. 

7.3.1.13. Sound propagation velocity 

Velocity of the acoustic signal propagating in the measured fluid. 

7.3.1.14. Fluid flow velocity 

Velocity of the fluid passing through the sensor flange. 

7.3.1.15. Start of the measurement period 

The calendar date, hour and minute when the measurement period commenced (the last resetting of the 
aggregate flow data). 

7.3.1.16. Duration of the measurement period 

The length of the period (in hours, minutes and seconds) from the measurement start (see section 7.3.1.15) 
during which the flow meter continuously performed the flow rate measurements. 

7.3.1.17. Duration of failure condition 

The total duration of the period(s) (in hours, minutes and seconds) from the measurement start during which the 
meter was energised but could not perform measurements due to a failure condition. 

7.3.1.18. Power failure period 

The total duration of the period(s) (in hours, minutes and seconds) from the measurement start during which the 
meter was not energised. 

7.3.1.19. Date 

The display shows the actual calendar date. 

7.3.1.20. Time 

The display shows the actual time of the day. 

7.3.2. Parameter setting mode 

When selecting the parameter setting mode (see section 7.3.1), the operator will be requested to enter a four-
digit password. 

7.3.2.1. Password 

 
The first display line will read 

PASSWORD 

 
 
and the first digit position on the second line will display 0. Depress push-button T3 repeatedly to increase the 
number by 1 at a time (after 9 will follow 0 again). Select the correct number at the first digit position and then 
depress push-button T2 to move to the second digit position and repeat the number setting procedure with push-
button T3. Progress to the third and fourth digit positions and enter the correct password – a combination of four 
numbers (see the schematic diagram in Fig. 7.3.3, page 32). 
Confirm the entry of the correct password by depressing push-button T4. Provided the password entered is 
correct, the system will proceed to the language selection block. In the case of an incorrect password the system 
will request a new password entry. After three consecutive entries of incorrect passwords the system will switch 
over to the data display mode and will not permit further entry into the parameter  
setting mode. A new attempt at the password entry is only possible after system de-energising and repeated 
switching on of the power supply. 

Summary of Contents for SONOELIS SE404X

Page 1: ...Manual Page 1 of 44 Ultrasonic flow meter SONOELIS SE404X SONOELIS SE406X ELIS PLZE a s Lucni 15 P O BOX 126 304 26 Plzen Czech Republic Phone 420 377517711 fax 420 377517722 Es90201K c Ultrasonic Fl...

Page 2: ...CONFIGURATION 19 7 2 THE STANDARD CONFIGURATION 19 7 2 1 Display data 20 7 2 1 1 Meter status information 20 7 2 1 2 Display of measured data 20 7 2 2 Review of the measured quantities 20 7 2 3 Review...

Page 3: ...ublic Phone 420 377517711 fax 420 377517722 Es90201K c 7 3 2 9 Meter zero setting 27 7 3 2 10 End of parameter setting 28 7 4 AUTOMATED METER TEST 28 8 SERVICE ACTIVITIES 36 8 2 WARRANTY SERVICES 37 8...

Page 4: ...een two ultrasonic transducers The flight times are measured for both directions of the signal wave propagation upstream and downstream whereby any asymmetry in the transducer positions are effectivel...

Page 5: ...fluid flow stabilisation piping sections at the meter input and output and generally increases the measurement accuracy In the case of a split version of the meter with separate sensor and the associa...

Page 6: ...he preset aggregate fluid mass level indication of measured values exceeding the preset temperature level meter failure indication The active state of the switching output can be selected either close...

Page 7: ...206 220 237 16 40 L mm 360 360 360 360 360 PN Nominal dim DN 100 DN 125 DN 150 DN 200 DN 250 DN 300 16 P mm 220 250 285 340 405 460 16 S mm 256 285 314 361 419 471 40 P mm 235 270 300 375 450 515 40 S...

Page 8: ...figuration 3 2 2 The compact version In the compact version the evaluation electronic unit is mounted directly onto the meter sensor UC 3 0 or UC 5 0 see section 3 2 1 1 above instead of the sensor te...

Page 9: ...ensure that no unauthorised modification of the meter functions or readings may take place The official certification and meter protection consist of Stick on labels with official seals of the respons...

Page 10: ...3 5 5 0 9 0 14 20 qNEC m3 h 0 07 0 09 0 14 0 22 0 37 0 6 0 9 1 2 2 2 3 5 5 Flow meter model SE406X DN 32 40 50 65 80 100 125 150 200 250 300 qs m3 h 20 32 50 80 150 240 350 500 900 1400 2000 qp m3 h 1...

Page 11: ...17 mm width 85 mm 1 5 kg 100 250 V 50 60 Hz 3 V Li battery Lifetime 5 years 6 VA T 250 mA 250 V automated disconnection from power supply in the TN S network Evaluation electronic unit compact version...

Page 12: ...selected sensor size can be re set in the COMFORT configuration using the customer control push button panel The graph below shows the pressure loss values at various flow rates for particular sensor...

Page 13: ...he ultrasonic flow meter shall be applied installed in observance of certain rules concerning meter placement in the fluid piping so as to ensure that the measurement accuracy complies with the meter...

Page 14: ...the sensor at all times Flow sensor location at the bottom pocket of the fluid piping If a flow sensor is to be installed in a vertical section of the fluid piping the fluid flow direction shall alway...

Page 15: ...and air bubbles either coming into the fluid through leaks in the piping or originating by the cavitation process in the sensor or other piping components If cavitation is suspected to appear in the s...

Page 16: ...ble conductors shall be soldered and the soldered joint protected against environmental and mechanical stresses by a suitable installation box All cables shall be led outside the thermal insulation la...

Page 17: ...C 5 0 W6 24V AC 0 1A RS 485 SWITCHING OUTPUT TRANSDUCER 1 TRANSDUCER 2 TRANSDUCER 3 IMPULSE OUTPUT OUTPUT CURRENT TRANSDUCER 4 50 60Hz Line power supply 100V 250V ON ON 2 FREQUENCY OUTPUT OFF 1 0 4 S2...

Page 18: ...ted to a relay coil in series with an external AC power source and so provide for indication of the flow direction or other selected parameter status Thermometer Pt 100 if used shall be connected to t...

Page 19: ...dges in the flow channel The separate box with electronic circuits in the case of a split version of the meter shall be attached to a suitable vertical support plate by means of four bolts of diameter...

Page 20: ...f the electronic unit will usually be manifested by discontinued regular changes of the system status symbols on the display 7 2 1 2 Display of measured data Up to three measured quantities can be dis...

Page 21: ...3 Review of the measured quantity units Volume flow rate Mass flow rate Volume Mass m3 hour t hour 1 000 m3 1 000 t m3 min t min m3 t m3 s t s l kg l hour kg hour 1 000 bbl 1 000 tons l min kg min bb...

Page 22: ...s plant Push buttons T1 to T4 are provided with the following graphic symbols T1 T2 T3 T4 The push button control of the meter is shown in a schematic diagram in see Fig 7 3 1 page 31 The system can...

Page 23: ...ows the ratio in per cent of the measured volume flow rate to the specified maximum volume flow rate 7 3 1 3 Mass flow rate The mass flow rate can be measured and the measured data displayed only on c...

Page 24: ...ould not perform measurements due to a failure condition 7 3 1 18 Power failure period The total duration of the period s in hours minutes and seconds from the measurement start during which the meter...

Page 25: ...hanges are done using push button T1 confirmation of the new value by push button T4 The display will then read PARAMETER SET To leave the current parameter setting mode and proceed to another paramet...

Page 26: ...nd the associated measuring unit will then appear on the first line of the display and the second line will show the previously defined limit value with the exception of the date and time The limit va...

Page 27: ...urement period the date hour and minute of the same The system will respond with a confirming message Parameter Set 7 3 2 9 Meter zero setting Before leaving the manufacturing plant every flow meter i...

Page 28: ...ro flow rate Then switch off the power depress push button S1 resetting the aggregate volume and with S1 depressed switch on the power again Upon releasing S1 the display will read TEST SENSOR FULL De...

Page 29: ...ual number of the minute of the cleaning procedure is displayed Every fourth seconds one numeral is added the line will be filled up by 15 same numbers within 1 minute after elapsing this time the dis...

Page 30: ...f physically possible values the following message will appear on the display UTS RATE LIMITS ADJUSTMENT and the actual limits will automatically be re adjusted with respect to the measured value The...

Page 31: ...Design Assembly and Service Manual Page 31 of 44 Ultrasonic flow meter SONOELIS SE404X SONOELIS SE406X ELIS PLZE a s Lucni 15 P O BOX 126 304 26 Plzen Czech Republic Phone 420 377517711 fax 420 377517...

Page 32: ...Design Assembly and Service Manual Page 32 of 44 Ultrasonic flow meter SONOELIS SE404X SONOELIS SE406X ELIS PLZE a s Lucni 15 P O BOX 126 304 26 Plzen Czech Republic Phone 420 377517711 fax 420 377517...

Page 33: ...Design Assembly and Service Manual Page 33 of 44 Ultrasonic flow meter SONOELIS SE404X SONOELIS SE406X ELIS PLZE a s Lucni 15 P O BOX 126 304 26 Plzen Czech Republic Phone 420 377517711 fax 420 377517...

Page 34: ...Design Assembly and Service Manual Page 34 of 44 Ultrasonic flow meter SONOELIS SE404X SONOELIS SE406X ELIS PLZE a s Lucni 15 P O BOX 126 304 26 Plzen Czech Republic Phone 420 377517711 fax 420 377517...

Page 35: ...Design Assembly and Service Manual Page 35 of 44 Ultrasonic flow meter SONOELIS SE404X SONOELIS SE406X ELIS PLZE a s Lucni 15 P O BOX 126 304 26 Plzen Czech Republic Phone 420 377517711 fax 420 377517...

Page 36: ...mbly and Service Manual Page 36 of 44 Ultrasonic flow meter SONOELIS SE404X SONOELIS SE406X ELIS PLZE a s Lucni 15 P O BOX 126 304 26 Plzen Czech Republic Phone 420 377517711 fax 420 377517722 Es90201...

Page 37: ...mmercial meter the parameters of a repaired product shall be verified at a duly authorised metrological centre 8 2 Post warranty services The post warranty services are understood to include any repai...

Page 38: ...on and maintenance manuals and a statement on the product compliance with the respective standards 13 WARRANTY CONDITIONS Unless agreed otherwise between the manufacturer and the customer the warranty...

Page 39: ...0 120 0 6 125 175 0 7 150 250 0 8 200 450 0 9 250 700 1 0 300 1000 1 1 non standard X X SN EN 1092 1 1 ANSI B 16 5 2 BS 4504 3 JIS B2210 4 non standard X carbon steel stainless steel painted of polyur...

Page 40: ...on standard X X FLOW METER SETTINGS one directional volume 1 one directional mass 2 bi directional volume 3 bi directional mass 4 non standard X standard qNEC according to manual 1 standard 10 mm s 2...

Page 41: ...on exceeding switch off 0 9 mass level on exceeding switch on 1 0 mass level on exceeding switch off 1 1 temperature level on exceeding switch on 1 2 temperature level on exceeding switch off 1 3 non...

Page 42: ...8 ton min 0 9 ton s 1 0 lb h 1 1 lb min 1 2 lb s 1 3 qp 1 4 non standard X X mass is not displayes 1 t 2 t 10 3 3 kg 4 ton US ton 5 lb libra 6 non standard X C 1 F 2 m s 1 ft s 2 Czech 1 English 2 Ge...

Page 43: ...n in 5 points with calibration certificate 5 standard calibration in 9 points without calibration certificate 6 standard calibration in 9 points with calibration certificate 7 standard calibration for...

Page 44: ...SONOELIS SE406X ELIS PLZE a s Lucni 15 P O BOX 126 304 26 Plzen Czech Republic Phone 420 377517711 fax 420 377517722 Es90201K c Address of manufacturer ELIS PLZE a s Lu n 15 P O BOX 126 304 26 Plze C...