ELIS PLZEŇ a. s.
Design, Assembly and Service Manual
Page 11 of 32
Electromagnetic flowmeter FLONET FN50xx.1
ELIS PLZEN a. s., Lucni 425/15, 301 00 Plzen, Czech Republic, Phone: +420/377 517 711, fax: +420/377 517 722
Es90716K
Values of flow Q
1
, Q
2
, Q
3
a Q
4
,
which apply to individual design variants and to dimensions, are written in the
table 2 below:
Table 2
Size DN
Over load
flow
Q4 [m3/h]
Permanent
flow
Q3 [m3/h]
Transition
flow
Q2 [m3/h]
Min. flow
Q1 [m3/h]
Insensitivity
[m3/h]
Q1 35%
Insensitivity
[m3/h]
Q1 50%
Ratio
Q3/Q1
15
6,5
5,2
0,0416
0,026
0,0091
0,013
200
20
12
9,6
0,0768
0,048
0,0168
0,024
25
18
14,4
0,1152
0,072
0,0252
0,036
32
30
24
0,192
0,12
0,042
0,06
40
45
36
0,288
0,18
0,063
0,09
50
72
57,6
0,4608
0,288
0,1008
0,144
65
120
96
0,768
0,48
0,168
0,24
80
180
144
1,152
0,72
0,252
0,36
100
280
224
1,79
1,12
0,392
0,56
125
430
344
2,75
1,72
0,602
0,86
150
650
520
4,16
2,6
0,91
1,3
200
1 150,0
920
7,36
4,6
1,61
2,3
250
1 800,0
1 440,0
11,52
7,2
2,52
3,6
300
2 520,0
2 016,0
16,13
10,08
3,528
5,04
350
3 500,0
2 800,0
22,4
14
4,9
7
400
4 500,0
3 600,0
28,8
18
6,3
9
500
7 200,0
5 760,0
46,08
28,8
10,08
14,4
600
10 000,0
8 000,0
64
40
14
20
700
14 000,0
11 200,0
89,6
56
19,6
28
800
18 000,0
14 400,0
115,2
72
25,2
36
Legend:
Q
4
overload (maximum) low (maximum)
Q
3
permanent (nominal) flow
Q
2
transition flow
Q
1
min. flow for specific accuracy