background image

 

13 

3.5 

Fundus observation with Hruby lens (optional) ------------------------- 

In  routine  applications,  only  the 
anterior  segment  of  the  vitreous 
body  can  be  examined  because  of 
the  refraction  effects  of  the  cornea 
and  crystalline  lens.  However,  with 
the Hruby  Lens, examination of the 
vitreous body is possible. 
 
Operation procedure: 

1)

 

The  pupil  should  be  dilated  for 
about 20 minutes. 

2)

 

Insert  the  Hruby  lens  guide  plate 
(15)  into  the  main  shaft  hole  of  the 
illumination  and  the  microscope 
arm.   

3)

 

Pull out the Hruby lens holder from 
one side of the headrest as shown in 
Fig.32. Move the Hruby lens holder 
toward  the  operator  so  that  it  can 
slide  freely  to  the  left  and  right 
below  the  chin-rest.    Insert  the 
lower  end  of  Hruby  lens  lever  into 
the groove on the guide plate. 

4)

 

Move  the  focus  of  the  illumination 
light  and  the  microscope  to  the 
patient’s eye. 

5)

 

Move  the  lever  to  locate  the  Hruby 
lens at the center of the vision field 
and near the patient’s eye, as shown 
in Fig.32. 

6)

 

Move  the  lever  to  focus  the  Hruby 
lens  at  the  fundus,  and  then  adjust 
the  slit  height  and  width  to  reduce 
the unnecessary interference light in 
the field-of-view.   

7)

 

To  examine  different  parts,  either 

turn  the  microscope  and  the 
illumination  arm  or  change  the 
patient’s  fixation  by  manipulating 
the fixation target. 

8)

 

If the long mirror interference in the 
examination, just replace it with the 
short mirror. 

9)

 

After examination, move the Hruby 
lens back to the original position on 
one side of the chin-rest.   

Attention:  before  moving  the 

Hruby  lens  to  the  right  and  left, 
first  have  the  patient’s  head 
move  away  from  the  chin-rest  to 
avoid  his  nose  touching  the 
Hruby lens.   

 

Fig.32 

 

Summary of Contents for ESL-7800

Page 1: ......

Page 2: ...rectly and firmly connected before using Ensure that the instrument is well grounded 3 Please pay attention to all the ratings of the electrical connecting terminal 4 Only use fuse according to the sp...

Page 3: ...Fig 1...

Page 4: ...n with Hruby lens optional 14 4 Maintenance 15 4 1 Replacing the illumination bulb 15 4 2 Replacing the reflecting mirror 15 4 3 Replacing the fuse 16 4 4 Replacing the chin rest paper 16 4 5 Adjustin...

Page 5: ...e arm represent the two arms angle When the 0 on the ring relates to the short mark at one side of the operator the right eyepiece may be blocked and the side of the patient the left eyepiece 12 Chin...

Page 6: ...ble One is for diopter adjustment to help the patient to see the target clearly while the other is an illuminated fixed spot 31 Reflecting Mirror Both long and short reflecting mirrors are provided Th...

Page 7: ...3 2 Assembly This section of the manual describes how to assemble ESL 7800 slit lamp All parts should be taken out with great care from the packing case before assembling 2 1 Components Fig2 1...

Page 8: ...Hruby Lens 1 I Hruby Lens Guide Plate 1 J Spare Illumination Bulb 1 K Chin rest Paper 1 L Focusing Test Rod 1 M Protection Cap 2 N Dust proof Cover 1 O Fixation Target 1 P Spare Long Reflecting Mirro...

Page 9: ...the input voltage and frequency of the instrument according to that of the mains 2 Assembling the Worktable F To attach the worktable on the motorized instrument table please screw off four M8x20mm b...

Page 10: ...tion arm with the screwdriver V Aim the assembly hole of the illumination arm at the brass shaft sleeve with care then put down let the shaft keeping close to the bottom surface well and simultaneousl...

Page 11: ...tput socket of the power box Insert the plug of the input power cable H into the input socket of the power box Remove the cable clips from the bottom of the work table with screw driver U and wrap the...

Page 12: ...onpart When the main power switch 8 of the power box is placed at I it turns on and O for turn off The main power switch should be set at the O position before connecting the input cable with the powe...

Page 13: ...perator is an ametropia he should adjust the eyepiece diopter One eyepiece with four short reticle lines witch is usually placed at the right side helps to focus accurately after accessories being att...

Page 14: ...target Slide the knob to adjust the diopter compensation within the range from 15D to 10D Fig 22 Fig 22 The fixation target with spot light is especially for the patient whose diopter exceeds 15D Fig...

Page 15: ...ircular beams of light are available at full aperture 9 8 5 3 1 0 2 dia respectively With a slit image the slit height can be changed continuously from 1 to 9mm which is indicated through the display...

Page 16: ...ig 29 Fig 29 6 Reflecting mirror Both short and long reflecting mirror are available in this instrument Use the long mirror in normal examination When the angle between the illumination part and the m...

Page 17: ...e groove on the guide plate 4 Move the focus of the illumination light and the microscope to the patient s eye 5 Move the lever to locate the Hruby lens at the center of the vision field and near the...

Page 18: ...otherwise the illumination may be uneven Fig 34 Attention The bulb is hot Place the lamp cap in the original position and rotate it clockwise and insert the connecting plugs Turn on the main power swi...

Page 19: ...n the paper is depleted pull upward two fixing pins of the chin rest and place a new package of paper then fix the fixing pins again Fig 38 Fig 38 4 5 Adjusting the tightness of the slit width knob If...

Page 20: ...rts such as chin rest bracket forehead rest belt with soft cloth dipped with soluble detergent or water sterilize with medicinal alcohol Fig 41 Attention Don t wipe with any corrosive detergent lest t...

Page 21: ...nge the fuse P16 The bulb is not assembled properly Assemble the bulb properly P15 The filter lever is in the middle position or in the position of gray filter Set the filter lever to the correct posi...

Page 22: ...25o C to 40 o C This instrument should be stored in a well ventilated room without corrosive gas where the relative humidity should be 10 to 80 and environment temperature 10o C to 40o C If the assem...

Page 23: ...to horizontal Slit inclination 5o 10o 15o 20o four steps Filter piece Heat absorbing gray red free blue Illumination bulb 6V20W halogen bulb Movement base Fore and back movement 90mm Left and right m...

Page 24: ......

Reviews: