8
O
O
p
p
e
e
r
r
a
a
t
t
i
i
o
o
n
n
M
M
a
a
n
n
u
u
a
a
l
l
2.6
Patient environment
The patient environment represents a space where there is a possibility of direct contact
between the patient or the operator and third person.
When another type of device is used in the patient environment, use a device that complies
with IEC 60601-1. If the devices that do not comply with IEC 60601-1 are used, it is necessary
to use an isolating transformer to power the device or to connect the devices to additional
protective grounding.
①
Patient environment (represented by dotted line, extending exactly 1.5 meters)
②
equipment
③④
Peripheral equipment(EN XXXXX and IEC XXXXX)
⑥
Power cord(
⑤
included protective earth)
①
②
③
④
⑤
⑥