Operator Manual
PURELAB Ultra Mk2 Version 3 11/08
Page 25
Step 6 - Fit Polishing Purification Pack
1. REMOVE a new Purification Pack from its packaging.
2. REMOVE the sealing plugs from inlet and outlet ports.
3. WET ‘O’ rings and SLIDE new Purification Pack into the
right hand position pushing upwards against pack
reader contacts.
4. POSITION
Purification Pack onto spigots.
5. PUSH into unit.
6. ENSURE that the Purification Pack is fully engaged and
dropped down past the retainers.
Step 7 - Acceptance of Purification Pack Installation
1. SWITCH
ON
power.
2. PRESS TICK
9
button to initiate programming of
Polishing Pack.
3. PRESS TICK
9
button to accept Polishing Pack
Installation date.
4. PRESS TICK
9
button to initiate programming of
Primary Pack.
5. PRESS TICK
9
button to accept Primary Pack
Installation date.
6. PRESS
TICK
9
button to calculate Replacement date.
7. PRESS
TICK
9
button to return to process off screen.
Step 8 - Purification Pack Rinse
1. PRESS PROCESS button. ALLOW calibration
routine to be completed and process screen to appear.
2. PRESS Dispense Controller. DIRECT dispense flow to
drain or collect in suitable container (±10 liters).
3. ROTATE Dispense Controller and INCREASE the
dispense rate to maximum.
Note:
The unit will start and initially during this stage
air/water will be purged from the unit via the
dispenser and the drain line.
4. ALLOW to dispense for 4 minutes.
5. RELEASE air from UMF (Analytic
and Ionic) by
carefully opening the air bleed located at the top of the
UMF Filter.
Note:
Place an absorbent cloth around the air bleed
to catch any water that is expelled with the air.
6. PRESS Dispense Controller to stop dispense.
Retainer
Spigots
Fit Purification Pack
UMF Bleed
UMF bleed
Purification Pack Installation Screens