Page 9
3.3 Technical
Specification
The Technical Specifications for the
PURELAB Ultra
are as follows:
Feedwater
*
Parameter Limits
Source - Originally from potable supply,
then pre-treated
Preferably reverse osmosis (RO) or filtered service deionisation (SDI) or distilled.
Note:
mixed bed or twin bed deionised supplies should be cation limited at exhaustion.
Fouling Index (max)
1 for all models. A 0.2 micron membrane prefilter is recommended for all non-RO feeds
Service Deionisation (SDI) - M
Ω
-cm 1
M
Ω
-cm minimum resistivity at exhaustion
Reverse Osmosis (RO) - µS/cm
Recommended < 30 µS/cm
Free Chlorine
0.05 ppm max.
TOC
Recommended 50 ppb max.
Carbon Dioxide
30 ppm max.
Silica
2 ppm max.
Particulates
Filtration down to 0.2 micron advisable to protect internal and/or point of use filters
Temperature
1-40°C - Recommended 10-15°C
Flowrate
(maximum requirement)
130 l/hr
Drain requirements (gravity fall with air gap).
Maximum during service
Up to 2 l/min
Feedwater Pressure
0.7 bar (10 psi), maximum, 0.07 bar (1 psi) minimum
* Contact technical support for advice on feedwaters outside the range listed.
Note:
Different system configurations are available for different feedwater sources. See system set up.
Dimensions
Height 490mm
(19.3")
Width 410mm
(16.2")
Depth 365mm
(14.4")
Weight
Scientific
14.0kg (30.8 lb)
Bioscience
14.5kg (32.0 lb)
Ionic
14.5kg (32.0 lb)
Analytic
15.0kg (33.1 lb)
Genetic
15.0kg (33.1 lb)
Connections
Inlet-quick connect
8mm (5/16") OD
Drain
8mm (5/16") OD
Flush - Bioscience/ Genetic
8mm (5/16") OD
Positioning
Wall, bench or under bench mounted
Environment
Clean dry indoor. Temp 5-40°C
Humidity max 80% non-condensing