ELGA PURELAB UHQ II & PS Service Manual
PURELAB UHQ II & PS Service issue 2 03/07
16
VALV12060 (230 volt units)
VALV11291 (110 volt units)
•
In UHQ PS replace with:
VALV30607 (230 volt units)
VALV30608 (110 volt units)
Note: Refill is inhibited whilst unit is in Process
mode.
Reservoir
overfills
•
Inlet solenoid
valve
•
Level control
faulty
•
Inlet valve stuck open. Replace valve as per
valve list above in “Reservoir not filling”
•
Check level control calibration –
see section 9.