Step 4 - Purity Alarm Setting
The following water purity alarm choices will be displayed.
Select the alarm setting required using the
button and
accept with the
9
button.
Note:
To reset any of the setup parameters,
restart from the set up MENU and
follow instructions from Step 1.
Setting Up Display and Replacement Timers
Turn the unit off at the power inlet module. To enter the
replacement timer set up, press the left hand touch pad button
and at the same time turn the power on.
A graphic of the pre-treatment cartridge and clock graphics are
displayed with hourly timer status.
CAUTION!
Before re-setting any of the cartridge
timers, ensure that the appropriate new
cartridges have been installed and
securely located correctly in the
PURELAB
Prima.
Step 1 - Pre-treatment Cartridge Replacement Timer
Setting this screen will cause the pre-treatment cartridge timer
to reset to the preset value of 4380 hours (6 months).
Press
9
to jump to the next consumable or
to initiate reset.
Press
9
to reset timer or press to abort reset.
Press
9
to jump to the next consumable.
Step 2 - Accessing the Process On Display Screens
The normal process screen will display newly installed SETUP
preferences showing the following process information:
•
Output water purity
•
Water
temperature
•
Process
mimic
•
Reservoir
level
•
Scroll and Print icons
You can scroll through the following display screens:
•
Pre-treatment cartridge replacement (hours remaining)
Step 3 - Report Printing
(Only if connected to LA 618)
To print a report press the icon.
Pre-treatment Cartridge Timer
Process On
Autorinse
Purity Alarm Setting
100 µS/cm
50 µS/cm
20 µS/cm