8. SANITIZATION
PROCEDURES
The
PURELAB Prima
unit has a sanitization port which allows the
sanitization agent to be introduced into the system in the form of a
tablet. Recommended maximum frequency of cleaning once per
month.
The unit is sanitized to destroy the bacteria within the pipework, and
the filters of the unit. Please read this entire section to become familiar
with the procedure before you start. Sanitization may be required in
the following circumstances:
•
Once a month to maintain low bacterial counts.
•
If the unit has not been used for a prolonged period of time.
Step 1 - Start Sanitization Cycle
1. ENSURE reservoir is less than 70% full of
treated water.
2. PRESS the PROCESS button to stop the
process.
3. TURN OFF the electrical supply.
4. RELIEVE residual pressure in the system by
waiting a few minutes after switching off the
power.
Step 2 - Insert Tablet into Sanitization Port
1. UNSCREW cap on sanitization port.
2. INSERT CT1 cleaning tablet.
3. REFIT cap on sanitization port, hand tight.
Step 3 - To Start the Sanitization Process
1. DISCONNECT outlet tube at the reservoir inlet
and re-direct to drain.
2. RESTORE the power.
3. PRESS the PROCESS button to start the
process.
4. ALLOW the sanitisation cycle to continue for 30
minutes.
5. PRESS the PROCESS button to stop the
process.
6. TURN electrical supply off.
Step 4 - Return to Normal Operation
1. RECONNECT the outlet tube to the reservoir.
2. TURN on power.
3. PRESS PROCESS button to return to normal
operation.
4. CHECK the system for leaks.