PURELAB Option-R 7/15
Operator Manual
ELGA
Page 32
PURELAB Option-R 7/15 Version 4 08/07
Step 7 - System Flushing
1. ENSURE the electrical supply is switched off.
2. DISCARD the majority of the reservoir contents
to drain, via the reservoir tap.
3. DISCONNECT outlet tube at the reservoir inlet
and direct to drain.
4. RESTORE the power supply.
5. PRESS the process button to start system
flushing. The system will operate automatically
during flushing and it is recommended that the
system is flushed overnight.
6. PRESS the process button to stop the system
flushing.
7. ENSURE the electrical power supply is switched
off.
Step 8 - Remove By-pass Block
1. PUSH and TILT by-pass block.
2. SLIDE out of unit.
3. PLACE in storage area.
Step 9 - Replace Ion-exchange Cartridge Pack
1. INSERT ion-exchange cartridge into the
compartment.
2. ENSURE the cartridge is locked in position.
3. CLOSE front door.
Step 10 - Return to Normal Operation
1. TURN on power.
2. RECONNECT the outlet tube to the reservoir.
3. PRESS PROCESS button to return to normal
operation.
4. CHECK the system for leaks.
Removing By-Pass Block
Replace Ion Exchange Cartridge