PURELAB Option-R 7/15
Operator Manual
ELGA
Page 30
PURELAB Option-R 7/15 Version 4 08/07
Step 4 - Remove Ion-exchange Cartridge Pack
1. REMOVE Ion-exchange cartridge.
2. PLACE cartridge safely to one side.
Step 5 - Fit Sanitization By-pass Block
1. WET ‘O’ rings on by-pass block.
2. SLIDE Sanitization by-pass block into unit.
3. ENSURE
Sanitization
by-pass block is locked in
place.
Step 6 - Start the Sanitization Process
1. DISCONNECT outlet tube at reservoir inlet and
re-direct to drain.
2. RESTORE the power.
3. PRESS the PROCESS button to start the re-
circulation process.
4. ALLOW the sanitization cycle to continue for 30
minutes.
5. PRESS the PROCESS button to stop the re-
circulation process.
6. TURN electrical supply off.
Step 7 - Remove By-pass Block
1. PUSH and TILT by-pass block.
2. SLIDE out of unit.
3. PLACE in storage area.
Step 8 - Replace Ion-exchange Cartridge Pack
1. INSERT ion-exchange cartridge into the
compartment.
2. ENSURE the cartridge is locked in position.
3. CLOSE front door.
Step 9 - Return to Normal Operation
1. TURN on power.
2. RECONNECT the outlet tube to the reservoir.
3. PRESS PROCESS button to return to normal
operation.
4. CHECK the system for leaks.
Removing Sanitization
By-Pass Block
Removing
Inserting
Fit Sanitization By-pass
Block