PURELAB Option-R 7/15
Operator Manual
ELGA
Page 6
PURELAB Option-R 7/15 Version 4 08/07
3.3 Technical
Specifications
The Technical Specifications for the
PURELAB Option-R
are as
follows:
Feedwater
PURELAB Option-R 7
PURELAB Option-R 15
Feedwater
Source Quality
Potable mains water supply
Potable mains water supply
Fouling Index-maximum
10
10
Total Dissolved Solids-maximum
1400
μ
S/cm 1400
μ
S/cm
Free Chlorine - maximum
0.5ppm
0.5ppm
Heavy Metals - maximum
0.05ppm
0.05ppm
Silica- maximum
30ppm
30ppm
Temperature
1 - 35°C
1 - 35°C
Flowrate
(Maximum requirement)
78 l/hr
85 l/hr
Drain Requirements (gravity fall with air gap). Maximum during
Service
70 l/hr
70 l/hr
Feedwater Pressure
Maximum - without internal boost pump
6.0 bar (90 psi)
6.0 bar (90 psi)
Minimum - without internal boost pump
4.0 bar (60 psi)
4.0 bar (60 psi)
Maximum - with internal boost pump
2.0 bar (30 psi)
2.0 bar (30 psi)
Minimum - with internal boost pump
Flooded Suction
Flooded Suction
Dimensions
Height
460mm (18.1")
460mm (18.1")
Width
550mm (21.7")
550mm (21.7")
Depth
270mm (10.6")
270mm (10.6")
Weight
With internal boost pump
20kg (44lb)
21kg (46lb)
Without internal boost pump
18kg (40lb)
19kg (42lb)
Connections
Inlet-quick connect
8mm (5/16") OD
8mm (5/16") OD
Outlet-quick connect
8mm (5/16") OD
8mm (5/16") OD
Drain RO-quick connect
8mm (5/16") OD
8mm (5/16") OD
Reservoir feed/return-quick connect
8mm (5/16") OD
8mm (5/16") OD
Positioning
Wall, bench or under bench
mounted.
Wall, bench or under bench
mounted.
Environment
Clean dry indoor.
Temp 5 - 40°C.
Humidity max 80%
non-condensing.
Clean dry indoor.
Temp 5 - 40°C.
Humidity max 80%
non-condensing.