![ELGA PURELAB Option-R 7 Operator'S Manual Download Page 20](http://html1.mh-extra.com/html/elga/purelab-option-r-7/purelab-option-r-7_operators-manual_2388426020.webp)
ELGA
PURELAB Option-R 7/15
Operator Manual
PURELAB Option-R 7/15 Version 4 08/07
Page
15
Step 4 - Positioning Dispense Tap
The Dispense Tap may be fitted at a high level or a low level on
the
PURELAB Option-R
unit, to allow for easier access when
wall mounted.
To alter the location of the Dispense Tap:
1. Switch
the
PURELAB Option-R
off at the power
switch at the top left hand side of the unit.
2. Ensure the water supply is turned off.
3. Dissipate pressure by opening the dispense tap.
4. Open the front door.
5. Remove the LC141 cartridge.
(See section 7.2).
6. Unscrew the two panel fixing screws located on
the right hand side of the door opening.
7. Remove side panel by sliding the panel to the
rear. Once unhooked, remove.
8. Remove screws holding dispense tap manifold.
9. Locate manifold in alternative position.
10. Replace panel.
11. Replace LC141 cartridge.
(See section 7.2).
12. Switch electrical power on.
13. Switch water supply on.
14. Press the process button.
15. Ensure correct operation of dispense tap.
Panel
fixing
screws
Slide back,
unhook and
remove side
panel
Manifold
screws
Replacing side panel
Panel
fixing
screws
Manifold
screws
Replace
side panel
Positioning the Dispense Tap
Summary of Contents for PURELAB Option-R 7
Page 1: ...PURELAB Option R 7 15 Operator Manual...