7.1 Replacing the LC140 Pre-treatment Cartridge
The pre-treatment cartridge should be replaced when indicated by the
change reminder.
Step 1 - Switch Unit Off
1. SWITCH the
PURELAB Option-R
off at the
power switch at the top left hand side of the unit.
2. ENSURE pressure has dissipated from the unit
by opening the dispense tap.
Step 2 - Remove Pre-treatment Cartridge
1. OPEN front door.
2. IDENTIFY the pre-treatment cartridge (LC140).
3. REMOVE the reducing fitting from the elbow at
the bottom of the cartridge, by pushing back the
retaining collet on the push fit connector and
withdrawing the reducer.
4. REMOVE the reducing fitting from the elbow at
the top of the cartridge.
5. REMOVE exhausted cartridge from retaining
clips and discard.
Note: The consumable is non-hazardous.
Dispose of as ordinary waste,
observing all local and national
regulations.
Step 3 - Replacing the Pre-treatment Cartridge
1. UNPACK new cartridge and remove the two
protective transit plugs sealing the inlet and outlet
connection.
2. SECURE the new cartridge into its retaining clips
ensuring the cartridge is the correct way up.
3. REFIT the inlet tubing into the bottom of the
cartridge by pushing the reducer into the elbow
connector until locked and held by the retaining
collet.
4. REFIT the outlet tubing at the top of the cartridge.
5. RESET pre-treatment cartridge timer.
(See
Section 5.4 - Setting Up Display and
Replacement Timer).
6. PRESS the PROCESS button to start the unit.
7. CHECK the unit for leaks and close front door.
Pre-treatment
cartridge
Location of Pre-treatment
Cartridge