PURELAB flex
Section 4 - Maintenance
Issue 4
Page
32
of
45
Calibration
The main power will be on while performing the calibration procedure,
please use caution
It is good practice to calibrate the system every 12 months. This must be carried out by an ELGA
trained service technician using the correct substitution (sub) boxes.
Sub boxes and adaptors required for the calibration of the PURELAB flex:
Sub Box 07
BOXS31509C
Sub Box 14
BOXS39060C
Sub Box 15
BOXS39061C
Adaptors LA738
These are included with Sub
boxes 14 & 15, You do not
need to purchase these
separately.
These substitution boxes must be calibrated annually by ELGA, check the calibration label before use to
make sure they are still in date. Please contact ELGA technical support for more information.
Connecting the calibration boxes
Sub box Adaptor Connection to the PCB
All SUB boxes have a square 4-position
connector and require an adaptor to be
able to connect to the PCB. These
adaptors are available from ELGA, part
number LA738
The PCB position for connecting the
quality SUB box adaptor is „JH2‟.
The PCB position for connecting the
temperature SUB box adaptor is „JH3‟.
!
JH2
JH3