Electrical Parameters
Most ELGA products are powered from a mains electrical supply, this could be up to 240V. Please
ensure that when performing maintenance on these units that the power supply is isolated whenever
possible before any work is carried out.
Please use caution when working on systems connected to a main
power supply. If in doubt do not atempt any work on a product unless
the power supply is isolated.
The electrical parameters detailed in this section of the manual are to be used as a guide to help assist
with general maintenance and troubleshooting. The values below can be measured using a multimeter.
Only the power supply supplied with the product should be used. Failure to use the correct power
supply will invalidate all regulatory compliance and warranty claims.
Electrical Supply
nominal
variance
AC Voltage Input to power supply
240 vac
±10%
100 vac
±10%
DC Voltage output from power supply
24 vdc