5.
MAINTENANCE
An approved supplier or distributor should carry out any maintenance work not included
in this manual.
Note:
Disposal of all end-of-life consumable items should be in accordance with local
statutory regulations.
5.1
Replacing Ultraviolet lamp (LC210)
The Ultraviolet (UV) lamp should be replaced when;
UV lamp failure occurs.
UV lamp efficiency decreases affecting water purity.
If either occurs you’ll be prompted to change the UV Lamp.
WARNING!
UV-C RADIATION IS HARMFUL TO THE EYES AND
SKIN. UV LAMP SHOULD ONLY BE OPERATED IN
THE CHAMBER.
WARNING!
IT IS STRONGLY RECOMMENDED THAT DURING
THE HANDLING OF THE UV LAMP CUT-RESISTANT
GLOVES ARE WORN.
MERCURY HAZARD, DO NOT BREAK. THE UV LAMP
CONTAINS A SMALL AMOUNT OF MERCURY. MAY
CAUSE REDNESS OR IRRITATION AS A RESULT OF
CONTACT WITH SKIN AND/OR EYES.
Step 1 – Switch Unit Off
Step 1 – Switch Unit Off
ISOLATE the electrical supply where the PURELAB flex power cable is connected
to the mains electrical supply.
TURN OFF the water supply.
OPEN the left hand access door.