Step 2 – Remove UV Lamp from PURELAB flex.
DISCONNECT the lamp connector fitted to the top of the UV lamp.
LOOSEN the 2 securing screws at the top of the UV housing.
LIFT and Remove the UV lamp from the UV housing.
DISCARD the used UV lamp in accordance with local regulations.
CAUTION!
Hold on to the UV lamp carefully to ensure it
doesn’t break.
Step 3 - Replace the UV lamp
UNPACK new UV lamp.
CAUTION!
Take care not to touch the surface of the glass.
Ideally wear gloves, handle with soft cloth and
wipe the surface with alcohol before fitting into
the housing.
SLIDE the new UV lamp into the UV housing.
CONNECT the lamp connector to the end of the UV making sure the connector is
correctly aligned with the pins.
TIGHTEN the 2 securing screws at the top of the UV housing.
CLOSE the left hand access door.
Step 4 – Turn Unit On
TURN ON the water supply.
SWITCH ON the mains electrical supply.