PURELAB flex 1/2 – US Operators Manual
ELGA
Page 12
PURELAB flex 1/2 - US MANU40562 Version 02 - 10/15
CONNECTIONS
Inlet
8 mm (5/16) OD tube
ELECTRICAL REQUIREMENTS
Mains input
100-240V ac, 50-60 Hz
System control voltage (not including
pumps and UV)
24 V dc
Power consumption (peak demand)
120 VA
PRODUCT WATER SPECIFICATION
Model
PURELAB flex
PURELAB flex
(with Purification
Pack)
PURELAB flex
(with UV,
Purification Pack
and TOC)
Delivery flow rate - maximum
Up to 2l/min
Up to 2l/min
Up to 2l/min
Inorganics (Resistivity @25°C)
As per
feedwater
18.2MΩ-cm
18.2MΩ-cm
Organics (TOC) – typically
Dependant on
feedwater
Dependant on
feedwater
<5ppb
2
Bacteria – Typical (when fitted with POU
filter)
<1 CFU/ml
<1 CFU/ml
<1 CFU/ml
Bacteria – Typical (when fitted with
Biofilter)
<1 CFU/10ml
<1 CFU/10ml
<1 CFU/10ml
Endotoxin (when fitted with Biofilter)
N/A
<0.001 EU/ml
<0.001 EU/ml
DNase (when fitted with Biofilter)
N/A
N/A
<20 pg/ml
RNase (when fitted with Biofilter)
N/A
N/A
0.002 ng/ml
The PURELAB flex can be upgraded to PURELAB flex with Purification pack, but cannot be
upgraded to PURELAB flex with UV, Purification Pack and TOC.
The PURELAB flex with Purification Pack cannot be upgraded to PURELAB flex with UV,
Purification Pack and TOC.
2
Dependent on feedwater.
As part of our policy of continual improvement we reserve the right to alter the
specifications given in this document.