ISL55180 EVM Getting Started
Rev A04: 11/28/2012
Copyright
Elevate Semiconductor Corporation 2012
Page 11 of 20
2.3.2 FV/MI
Configuration
Figure 3 illustrates the recommended configuration for FV/MI (Force Voltage/Measure Current). After the
configuration is completed, use the
Europa->Channel#0->Levels
dialog box to change the ISL55180
output levels. Use the
Europa Loadboard
dialog box to connect other channels to the TEST_NODE.
Both MI_MONITOR and MONITOR are configured to output Channel#0 MI-S
If using an external source measurement unit (SMU), the SMU should be configured in the opposite mode
as ISL55180.
ISL55180
SMU
FV/MI FI/MV
FI/MV FV/MI
Figure 3: ISL55180 EVM FV/MI Simplified Block Diagram
Motherboard
Europa
Load Board
PC
J2
(FVMI
)
Parallel
Cable
EVM
Power
Supply
ISL55180
DMM
or
SMU
MONITOR
MONITOR
FORCE_0
SENSE_0
FORCE_0
VCCO
VEE
+20V (BN1)
-15V (BN2)
+5V (BN3)
GND (BN4)
LEDs
SENSE_0
TEST_NODE
Europa
Power
Supply
FORCE_7
SENSE_7
FORCE_7
SENSE_7
MI_MONITOR
MI_MONITOR
GND