elco ELK 38 Operating Instructions Manual Download Page 4

ELCO S.r.l. - ELK 38 - OPERATING INSTRUCTIONS - Vr. 02 - ISTR 06261 - PAG. 4

 

Connect  the  instrument  as  far  away  as  possible  from  sources  of 
electromagnetic disturbances such as motors, power relays, relays, 
solenoid valves, etc. 
 
3.3 - ELECTRICAL CONNECTION 
Carry  out  the  electrical  wiring  by  connecting  only  one  wire  to  each 
terminal,  according  to  the  following  diagram,  checking  that  the 
power  supply  is  the  same  as  that  indicated  on  the  instrument  and 
that  the  load  current  absorption  is  no  higher  than  the  maximum 
electricity current permitted.  
As  the  instrument  is  built-in  equipment  with  permanent  connection 
inside  housing,  it  is  not  equipped  with  either  switches  or  internal 
devices  to  protect  against  overload  of  current:  the  installation  will 
include  a  two-phase  circuit-breaker,  placed  as  near  as  possible  to 
the instrument, and located in a position that can easily be reached 
by  the  user  and  marked  as  instrument  disconnecting  device  which 
interrupts the power supply to the equipment. 
It  is  also  recommended  that  all  the  electrical  circuits  connected  to 
the  instrument  must  be  protect  properly,  using  devices  (ex.  fuses) 
proportionate to the circulating currents.  
It  is  strongly  recommended  that  cables  with  proper  insulation, 
according to the working voltages and  temperatures, be used.  
Furthermore,  the  input  cable  of  the  probe  has  to  be  kept  separate 
from line voltage wiring. If the input cable of the probe is screened, it 
has to be connected to the ground with only one side.  
Whether the instrument is 12 V version it’s recommended to use an 
external  transformer  TCTR,  or  with  equivalent  features,  and  to  use 
only  one  transformer  for  each  instrument  because  there  is  no 
insulation between supply and input. 
We  recommend  that  a  check  should  be  made  that  the  parameters 
are those desired and that the application functions correctly before 
connecting  the  outputs  to  the  actuators  so  as  to  avoid 
malfunctioning  that  may  cause  irregularities  in  the  plant  that  could 
cause damage to people, things or animals. 
ELCO  S.r.l.    and  its  legal  representatives  do  not  assume  any 
responsibility  for  any  damage  to  people,  things  or  animals 
deriving  from  violation,  wrong  or  improper  use  or  in  any  case 
not in compliance with the instrument’s features. 
 
3.4 - ELECTRICAL WIRING DIAGRAM 

 
 

4 - FUNCTIONS 

 
4.1 - MEASURING AND VISUALIZATION 
All  the  parameters  referring  measurements  are  contained  in  the 
group 

]

InP”. 

Depending on the model required the input accept: 
C:  Thermocouples  temperature  probes  (J,K,S  and  ELCO  IRS 
Infrared  sensors),  mV  signals  (0..50/60  mV,  12..60  mV), 
Thermoresistances PT100. 
E  :  Thermocouples  temperature  probes  (J,K,S  and  ELCO  IRS 
Infrared  sensors),  mV  signals  (0..50/60  mV,  12..60  mV), 
Thermistors PTC and NTC. 

I : normalized analogue signals 0/4..20 mA 
V : normalized analogue signals 0..1 V, 0/1..5 V, 0/2..10 V 
Depending on the model, using par. “SEnS”, it’s possible to select 
the type of input probe, which can be : 
-  for  thermocouples  J  (J),  K  (CrAL),  S  (S)  or  for  infrared  sensors 
serie ELCO IRTC1 with linearization J (Ir.J) or K (Ir.CA) 
- for thermoresistances Pt100 IEC (Pt1) or thermistors PTC KTY81-
121 (Ptc) or NTC 103AT-2 (ntc) 
-  for  normalised  signals  in  current  0..20  mA  (0.20)  or  4..20  mA 
(4.20) 
-  for  normalised  signals  in  tension  0..1  V  (0.1),  0..5  V  (0.5),  1..5  V 
(1.5), 0..10 V (0.10) or 2..10 V (2.10). 
- for normalised signals in tension 0..50 mV (0.50), 0..60 mV (0.60), 
12..60 mV (12.60). 
We  recommend  to  switch  on  and  off  the  instrument  when  these 
parameters are modified, in order to obtain a correct measuring. 
For  the  instruments  with  input  for  temperature  probes  (tc,  rtd)  it’s 
possible  to  select,  through  par.  “Unit”,  the  unit  of  measurement  
(°C,  °F)  and,  through  par.  “dP”  (Pt100,  PTC  and  NTC  only)  the 
desired resolution (0=1°; 1=0,1°). 
Instead,  with  regards  to  the  instruments  with  normalised  analogue 
input  signals,  it  is  first  necessary  to  program  the  desired  resolution 
on  par.  “dP”    (0=1;  1=0,1;  2=0,01;  3=0,001)  and  then,  on  par. 
"SSC", the value that the instrument must visualise at the beginning 
of  the  scale  (0/4  mA,  0/12  mV,  0/1  V  o 0/2 V) and, on par. "FSC", 
the value that the instrument must visualise at the end of the scale 
(20 mA, 50 mV, 60 mV, 5 V or 10 V). 
The instrument allows for measuring calibration, which may be used 
to  recalibrate  the  instrument  according  to  application  needs,  by 
using par. “OFSt” and “rot”. 
Programming  par.  “rot”=1,000,  in  par.  “OFSt”  it  is  possible  to  set  a 
positive or negative offset that is simply added to the value read by 
the  probe  before  visualisation,  which  remains  constant  for  all  the 
measurements. 
If instead, it is desired that the offset set should not be constant for 
all  the  measurements,  it  is  possible  to  operate  the  calibration  on  
any two points.  
In  this  case,  in  order  to  decide  which  values  to  program  on  par. 
“OFSt” and “rot”, the following formulae must be applied : 

“rot” = (D2-D1) / (M2-M1)  

“OFSt” = D2 - (“rot” x M2) 

where: 
M1 =measured value 1 
D1 = visualisation value when the instrument measures M1 
M2 =measured value 2 
D2 = visualisation value when the instrument measures M2 
It then follows that the instrument will visualise : 

DV = MV x “rot” + “OFSt” 

where:  DV = visualised value  

MV= measured value 

 
Example  1:  It  is  desired  that  the  instrument  visualises  the  value 
effectively  measured  at  20°  but  that,  at  200°,  it  v isualises  a  value 
lower than 10° (190°).     
Therefore :  M1=20 ;  D1=20 ;  M2=200 ;  D2=190 
“rot” = (190 - 20) / (200 - 20) = 0,944 
“OFSt” = 190 - (0,944 x 200) = 1,2 
Example 2: It is desired that the instrument visualises 10° whilst the 
value actually measured is 0°, but, at 500° it visu alises a 50° higher 
value (550°). 
Therefore :  M1=0 ;  D1=10 ;  M2=500 ;  D2=550 
“rot” = (550 - 10) / (500 - 0) = 1,08 
“OFSt” = 550 - (1,08 x 500) = 10 
 
By  using  par.  “FiL”  it  is  possible  to  program  time  constant  of  the  
software filter for the input value measured, in order to reduce noise 
sensitivity (increasing the time of reading). 
In case of measurement error, the instrument supplies the power as 
programmed on par.  “OPE”. 
This  power  will  be  calculated  according  to  cycle  time  programmed 
for  the  PID  controller,  while  for  the  ON/OFF  controllers  the  cycle

 

time  is  automatically  considered  to  be  equal  to  20  sec.  (e.g.  In  the 
event of probe error with ON/OFF control and “OPE”=50, the control 
output will be activated for 10 sec., then it will be deactivated for 10 
sec. and so on until the measurement error remains.). 

ELK

Summary of Contents for ELK 38

Page 1: ...ACTION PID CONTROL 4 6 DOUBLE ACTION PID CONTROL 4 7 AUTO TUNING AND SELF TUNING FUNCTIONS 4 8 REACHING OF SET POINT AT CONTROLLED SPEED AND AUTOMATIC COMMUTATION BETWEEN TWO SET POINTS 4 9 SOFT STAR...

Page 2: ...er two seconds in the same condition the changing speed increases in order to allow the desired value to be reached rapidly Once the desired value has been reached by pushing key P it is possible to e...

Page 3: ...of the controller During automatic control it is possible to visualize the control power on the display by pushing key UP The range of the power values goes from H100 100 of the output power with rev...

Page 4: ...nearization J Ir J or K Ir CA for thermoresistances Pt100 IEC Pt1 or thermistors PTC KTY81 121 Ptc or NTC 103AT 2 ntc for normalised signals in current 0 20 mA 0 20 or 4 20 mA 4 20 for normalised sign...

Page 5: ...value SP HSEt Vice versa in case of direct action or cooling Func CooL it deactivates the output when the process value reaches SP HSEt in case of symmetrical hysteresis or SP in case of asymmetrical...

Page 6: ...control monitoring and the continuous calculation of the parameters during control Both functions automatically calculate the following parameters Pb Proportional Band tcr1 Cycle time of the output 1r...

Page 7: ...on only works through PID control and allows the limitation of control power when the instrument is switched on for a programmable period of time This is useful when the actuator driven by the instrum...

Page 8: ...ument is in alarm condition the alarm is not activated It will be activated only when the process value is in non alarm conditions and then back in alarm conditions ALARM DELAY the alarm output may be...

Page 9: ...CONFIGURATION BY KEY01 The instrument is equipped with a connector that allows the transfer from and toward the instrument of the functioning parameters through the device ELCO KEY01 with 5 poles conn...

Page 10: ...arm AL1L 9999 9999 27 HAL1 Alarm AL1 hysteresis OFF 9999 1 28 AL1d Activation delay of alarm AL1 OFF 9999 sec OFF 29 AL1i Alarm AL1 activation in case of measuring error no yES no Group LbA parameters...

Page 11: ...defines the error conditions of the input allowing the instrument to give the power programmed on par OPE as output The possibilities are Or the condition occurs in case of over range or probe breakag...

Page 12: ...with Func HEAt than SP SP 2 or higher with Func CooL than SP SP 2 OFF Autotuning disabled When it s occurring an Autotuning cycle led AT blinks SELF SELF TUNING ENABLE Parameter used to enable yES or...

Page 13: ...y to repeat the auto tuning LbA Loop control interrupted Loop break alarm Check the working of probe and actuator and swap the instrument to rEG control ErEP Possible anomaly of the EEPROM memory Push...

Page 14: ...C 58 0 230 0 F 0 20 mA SEnS 0 20 1999 9999 199 9 999 9 19 99 99 99 1 999 9 999 4 20 mA SEnS 4 20 1999 9999 199 9 999 9 19 99 99 99 1 999 9 999 0 50 mV SEnS 0 50 1999 9999 199 9 999 9 19 99 99 99 1 99...

Reviews: