PRODUCT MAINTENANCE
02/2020
43
NP-DK50 PLUS, 2V-A-3_02-2020-MD
17.10. Procedure for connecting a compressor disconnected from the cabinet
Prior to any maintenance or repair work, switch off the compressor and
disconnect it from the mains (pull out the mains plug).
The compressor (without cabinet) requires that the jumper is always mounted in the terminal strip
for correct operation (Fig. 21, Fig. 22, Pos. B). It then replaces the switch on the cabinet.
If there is no jumper in the terminal strip, the compressor will not work!
If the cord between the compressor and the cabinet becomes disconnected (by pulling the mains
plug) and the compressor is removed from the cabinet, the compressor will not be functional.
Therefore, it is necessary to first create a connection to the terminal strip with a jumper (replacing
the circuit breaker function) (Fig. 21, Fig. 22).
PROCESS:
Jumper installation
(A-B process):
Disconnect the product from the mains by pulling the plug out of the socket.
Remove the cover on the electrical panel (on the compressor).
The jumper is not in the terminal strip
– A.
Insert the jumper into the terminal strip
– B.
Reinstall the cover on the electrical panel.
Connect the device to the electrical mains.
Activate the compressor by turning on the switch on the pressure switch.
Compressor - when operating outside the cabinet, the electrical connection with an installed jumper
must
be created (Fig. 21, Fig. 22, Pos. B).
Summary of Contents for DK50 2V S
Page 1: ...DK50 PLUS M DK50 2V M User manual Benutzerhandbuch N vod na pou itie...
Page 2: ......
Page 4: ...CONTENTS 5 INHALT 49 OBSAH 94...
Page 15: ...PRODUCT DESCRIPTION 02 2020 15 NP DK50 PLUS 2V A 3_02 2020 MD Fig 3 Cabinet...
Page 25: ...INSTALLATION 02 2020 25 NP DK50 PLUS 2V A 3_02 2020 MD Fig 6 Changing door direction...
Page 44: ...PRODUCT MAINTENANCE NP DK50 PLUS 2V A 3_02 2020 MD 44 02 2020 Fig 21 230V A B...
Page 59: ...PRODUKTBESCHREIBUNG 02 2020 59 NP DK50 PLUS 2V A 3_02 2020 MD Abb 3 Kompressorgeh use...
Page 69: ...INSTALLATION 02 2020 69 NP DK50 PLUS 2V A 3_02 2020 MD Abb 6 ndern der T r ffnungsrichtung...
Page 89: ...PRODUKTWARTUNG 02 2020 89 NP DK50 PLUS 2V A 3_02 2020 MD Abb 21 230V A B...
Page 104: ...POPIS V ROBKU NP DK50 PLUS 2V A 3_02 2020 MD 104 02 2020 Obr 3 Skrinka...
Page 114: ...IN TAL CIA NP DK50 PLUS 2V A 3_02 2020 MD 114 02 2020 Obr 6 Zmena otv rania dver...
Page 133: ...DR BA V ROBKU 02 2020 133 NP DK50 PLUS 2V A 3_02 2020 MD Obr 21 230V A B...
Page 141: ......
Page 142: ......
Page 143: ......