![EKOM DK50 2V/M MOBILE User Manual Download Page 11](http://html1.mh-extra.com/html/ekom/dk50-2v-m-mobile/dk50-2v-m-mobile_user-manual_2374248011.webp)
PRODUCT DESCRIPTION
10/2022
11
NP-DK50 2V-A-4_10-2022-AD
The regulator must be selected according to the application to the filter set, or separately The
regulator shall ensure constant pressure at the outlet.
Type
Use
Article number
Regulator complete
(to the filter assembly)
DK50 2V/M
to order
REG11 (separate)
447000001-077
Filter set brackets
A suitable bracket must be ordered for every filter set.
Equipotential bonding socket
The socket provides protective bonding.
Type
Use
Name
Article no.
POAG-KBT6-EC
DK50 2VS/M
Mains plug
connector
033400075-000
FLEXI-S/POAG-HK6
Conductor (1 m)
034110083-000
8. PRODUCT FUNCTION
8.1.
Compressor with adsorption dryer
(Fig. 1) The air pump (1) draws in air through the inlet filter (7) and supplies compressed air through
the connecting hose into the adsorption dryer (8). From the inlet to the dryer module, the air is first
cooled in the integrated cooler (26) and then moves through the condensate separator (32), entering
the active chamber with adsorbent (27), where the air is dried. A portion of the air is fed into the
second, regenerated chamber, where this air is used to remove moisture from the adsorbent, after
which it is vented to the atmosphere. The active chamber switches on a cyclical basis. The dry air
then passes through the non-return valve (3) into the air tank (2). Air drying efficiency (the pressure
dew point) is determined by the dew point sensor located in the air tank.
It is not necessary to drain the air tank.
8.2.
Compressor in cabinet
(Fig. 3) The soundproof cabinet provides compact covering of the compressor, dampens noise
efficiently, providing sufficient exchange of cooling air. Its design makes it suitable for use in rooms
where low noise levels are required. The fan under the compressor pump provides cooling of the
compressor, it is in operation at the same time with compressor motor or after switching on the
temperature switch at temperature above 40 °C. After cooling the cabinet area approximately under
32°C, the fans turn off automatically.
The dryer display (34) is located on the front panel of the cabinet.
Type
Use
Article number
Compressor-mounted bracket
DK50 2V/M
to order
Summary of Contents for DK50 2V/M MOBILE
Page 1: ...N VOD NA POU ITIE SK USER MANUAL EN N VOD K POU IT CS BENUTZERHANDBUCH DE...
Page 2: ......
Page 4: ...CONTENTS 5 INHALT 62 OBSAH 120 OBSAH 176...
Page 15: ...PRODUCT DESCRIPTION 10 2022 15 NP DK50 2V A 4_10 2022 AD Fig 3 Cabinet...
Page 72: ...PRODUKTBESCHREIBUNG NP DK50 2V A 4_10 2022 AD 72 10 2022 Abb 3 Geh use...
Page 83: ...INSTALLATION 10 2022 83 NP DK50 2V A 4_10 2022 AD A B C...
Page 130: ...POPIS V ROBKU NP DK50 2V A 4_10 2022 AD 130 10 2022 Obr 3 Skrinka...
Page 142: ...IN TAL CIA NP DK50 2V A 4_10 2022 AD 142 10 2022 Obr 6 Pripojenie tlakovej hadice...
Page 186: ...POPIS V ROBKU NP DK50 2V A 4_10 2022 AD 186 10 2022 Obr 3 Sk ka...
Page 198: ...INSTALACE NP DK50 2V A 4_10 2022 AD 198 10 2022 Obr 6 P ipojen tlakov hadice...
Page 236: ......
Page 237: ......
Page 238: ......
Page 239: ......