INSTALLATION
CONDITIONS FOR USE
Compressors may only be installed and operating in dry, well-ventilated and clean environments
with environmental parameters that meet the requirements specified in Chapter 7, Technical
data. The compressor must be installed so that it is accessible at all times for operating and
maintenance. Please ensure that the label on the device is readily accessible.
The compressor must stand on a flat, sufficiently stable base (be aware of the weight of the
compressor, see Chapter 7, Technical data).
Compressors may not be operated outdoors or in otherwise wet or damp environments. Do not
use the compressor in the presence of explosive gases, dust or flammable liquids.
Before connecting the compressor to medical equipment, the supplier must confirm that it meets
all requirements for its use. Refer to the technical data of the product for this purpose. When a
unit is to be built-in, classification and evaluation of compatibility must be done by the
manufacturer or supplier of the product to be used.
Any other use or use outside this framework is not considered as intended use. The
manufacturer is not responsible for any damages arising from such use. The operator/user
assumes all risks.
Only a qualified professional can install the compressor and place it into operation
for the first time. This professional is obliged to train operating staff as to the use
and maintenance of the device. Installation and training of all operators shall be
confirmed by the installer’s signature on the certificate of installation.
Prior to installation, ensure that the compressor is free of all transport packaging
and stabilizers to avoid any risk of damage to the product.
Parts of the air pump may be hot and reach hazardous temperatures during
compressor operation and may pose a contact risk for operators or materials. Burn
or fire hazard! Caution! Hot surface!
Ambient operating conditions
Temperature:
+5°C to +40°C,
Max. relative humidity:
70%,
Max. absolute humidity:
15 g/m
3
.
Summary of Contents for DK-50 2x2V/110
Page 2: ......
Page 4: ...CONTENTS 1 INHALT 49 SOMMAIRE 97 ОГЛАВЛЕНИЕ 145 ...
Page 25: ...DK50 2V 50 DK50 2x2V 110 06 2020 21 NP DK50 2V 50 2x2V 110 13_06 2020 MD A 3B DK50 2x2V 110S ...
Page 73: ...DK50 2V 50 DK50 2x2V 110 06 2020 69 NP DK50 2V 50 2x2V 110 13_06 2020 MD A 3B DK50 2x2V 110S ...
Page 121: ...DK50 2V 50 DK50 2x2V 110 06 2020 117 NP DK50 2V 50 2x2V 110 13_06 2020 MD A 3B DK50 2x2V 110S ...
Page 169: ...DK50 2V 50 DK50 2x2V 110 06 2020 165 NP DK50 2V 50 2x2V 110 13_06 2020 MD A 3B DK50 2x2V 110S ...
Page 202: ...No text ...
Page 203: ...No text ...