To Use the Monitor for a Long Time

 

Quality control

 •

The display quality of monitors is affected by the quality level of input signals and the degradation of 

the product. Perform visual checks and periodic constancy tests to comply with medical standards / 

guidelines according to your application, and carry out calibration as necessary. Use of the RadiCS 

monitor quality control software enables you to perform high-level quality control that meets medical 

standards / guidelines.

 • It takes about 15 minutes (under our measurement conditions) for the monitor display to stabilize. 

Please wait 15 minutes or more after the monitor power has been turned on or the monitor 

has recovered from the power saving mode before performing various tests for quality control, 

calibration, or screen adjustment of the monitor.

 • We recommend that monitors be set to the recommended level or lower to reduce changes in 

luminosity caused by long-term use and maintain stable brightness.

 • To adjust measurement results of the integrated calibration sensor (Integrated Front Sensor) to 

those of an EIZO external sensor (UX1 or UX2 sensor) that is sold separately, perform correlation 

between the Integrated Front Sensor and the external sensor using RadiCS / RadiCS LE. Periodical 

correlation allows you to maintain the measurement accuracy of the Integrated Front Sensor at a 

level equivalent to that of the external sensor.

Attention

 • The display status of the monitor may change unexpectedly due to an operating error or unexpected setting 

change. Using the monitor with the control buttons locked is recommended after adjusting the screen of the 
monitor. For details on how to set, refer to the Installation Manual (on the CD-ROM).

 

Cleaning

Periodic cleaning is recommended to keep the monitor looking new and to prolong its operation 
lifetime.
Gently wipe off any dirt on the cabinet or panel surface with a soft cloth soaked in a small amount of 
water or one of the chemical solutions listed below or with ScreenCleaner (optional product sold 
separately).

Chemicals that may be used for cleaning

Material name

Product name

Ethanol

Ethanol

Isopropyl alcohol

Isopropyl alcohol

Alkyldiaminoethylglycine

Tego 51

Glutaral

Sterihyde

Glutaral

Cidex Plus28

Attention

 •

Do not use chemicals on a frequent basis. Chemicals such as alcohol and antiseptic solution may cause 
gloss variation, tarnishing, and fading of the cabinet or panel, and also quality deterioration of the image.

 • Never use any thinner, benzene, wax, and abrasive cleaner, which may damage the cabinet or panel.
 • Do not let chemicals come into direct contact with the monitor.

To Use the Monitor Comfortably

 • Staring at the monitor for a long time tires your eyes. Take a 10-minute rest every hour.

 • Look at the screen from a proper distance and from a proper angle. 

 

8

Notice for This Monitor

Summary of Contents for RadiForce RX1270

Page 1: ...yourself with safe and effective usage Please retain this manual for future reference For monitor adjustment and settings refer to the Installation Manual For the latest product information including...

Page 2: ...ted specifically for use in the region to which it was originally shipped If operated outside this region the product may not perform as stated in the specifications No part of this manual may be repr...

Page 3: ...tes Main Power Switch Press to turn the monitor s main power off Main Power Switch Press to turn the monitor s main power on Power button Press to turn the monitor s power on or off Alternating curren...

Page 4: ...ny form of transportation ships aircraft trains automobiles etc Do not place in dusty or humid environments Do not place in locations where water may be splashed on the screen bathrooms kitchens etc D...

Page 5: ...he unit securely Otherwise the unit may become detached resulting in injury and or equipment damage Before installation make sure that desks walls or any other installation surface has adequate mechan...

Page 6: ...t install the unit in a place with poor ventilation or inadequate space Do not use the unit laid down or upside down Blocking the ventilation slots prevents proper airflow and may result in fire elect...

Page 7: ...onitor displays continuously over a long period of time dark smudges or burn in may appear To maximize the life of the monitor we recommend the monitor be turned off periodically The backlight of the...

Page 8: ...uracy of the Integrated Front Sensor at a level equivalent to that of the external sensor Attention The display status of the monitor may change unexpectedly due to an operating error or unexpected se...

Page 9: ...Installing the Product 16 Installation Requirements 16 2 2 Connecting Cables 16 2 3 Attaching and Removing RadiLight Focus Spotlight 18 2 4 Turning On the Power 19 2 5 Adjusting the Screen Height and...

Page 10: ...etween monochrome and color parts of the same image at a pixel level and displays them respectively in optimal gradations High degree of freedom of layout This product has the PbyP Picture by Picture...

Page 11: ...he monitor allowing users to interpret images effectively even in a dark place RadiLight Focus spotlight RadiLight Focus is the removable spotlight This light illuminates the hands of a user when read...

Page 12: ...allows you to display or hide the PinP sub window and switch the PC used to operate USB devices at the same time For more information about the setup procedure refer to RadiCS RadiCS LE User s Manual...

Page 13: ...le reference procedures Readme txt file RadiCS LE monitor quality control software for Windows User s Manual Monitor Installation Manual RadiCS LE User s Manual Outline dimensions RadiCS LE RadiCS LE...

Page 14: ...ion on how to install and use RadiCS LE refer to RadiCS LE User s Manual on the CD ROM When using RadiCS LE connect the monitor to your PC using the supplied USB cable For more information about how t...

Page 15: ...in backlight function of the monitor This light illuminates the room indirectly from the back of the monitor 7 Stand The height and angle can be adjusted 8 USB downstream port Connect it to a USB dev...

Page 16: ...with the screen 2 2 Connecting Cables Attention Check that the monitor PC and peripherals are powered off When replacing the current monitor with this monitor refer to 4 2 Compatible Resolutions page...

Page 17: ...byP mode to display images from two PCs use of some quality control features such as calibration may be restricted For DisplayPort 1 DisplayPort 2 For One Cable PbyP Signal cable PP300 V14 Signal cabl...

Page 18: ...SB device USB connected peripheral device to the monitor USB cable UU300 Attention When connecting the monitor to a PC on which RadiCS RadiCS LE has been installed connect the cable to Remove the cap...

Page 19: ...When not using the monitor you can turn off the main power supply or disconnect the power plug so that the power is cut completely Note In order to maximize the monitor s lifespan by impeding brightn...

Page 20: ...the Cable Cover 1 Tidy the cables into the cable holder 2 Install the cable cover Example Installing the Cable Cover Right Attention Install the cover properly to ensure ventilation inside the monito...

Page 21: ...and brightness of LED lights are not consistent even among the products of the same model RadiLight Area may not sufficiently light up the room in the following environment If there is a relatively l...

Page 22: ...pecified by EIZO Turn off the main power and then turn it on again If the signal cable is connected to DisplayPort 1 try switching the DisplayPort version For details refer to the Installation Manual...

Page 23: ...kHz 175 kHz HDMI 31 kHz 160 kHz Vertical scan frequency 2 DisplayPort 29 Hz 61 Hz 720 400 69 Hz 71 Hz HDMI 59 Hz 61 Hz 720 400 69 Hz 71 Hz Frame Synchronization mode 29 5 Hz 30 5 Hz 59 Hz 61 Hz Dot c...

Page 24: ...4 The USB upstream port is not connected DP Power Save is set to On DisplayPort 1 is set to Version 1 2 DisplayPort 2 is set to Version 1 2 and no external load is connected 4 2 Compatible Resolutions...

Page 25: ...e optional accessories and information about the latest compatible graphics board refer to our web site www eizoglobal com Calibration Kit RadiCS UX2 Ver 5 0 4 or later RadiCS Version Up Kit Ver 5 0 4...

Page 26: ...that could influence limit or result in malfunction of the monitor Install the equipment in a controlled environment where such effects are avoided Classification of Equipment Type of protection agai...

Page 27: ...equipment If adjacent or stacked use is necessary the equipment or system should be observed to verify normal operation in the configuration in which it will be used When using a portable RF communic...

Page 28: ...atic discharge ESD IEC EN61000 4 2 8 kV contact discharge 15 kV air discharge 8 kV contact discharge 15 kV air discharge Floors should be wood concrete or ceramic tile If floors are covered with synth...

Page 29: ...ould be less than the compliance level in each frequency range b Interference may occur in the vicinity of equipment marked with the following symbol Note 1 UT is the a c mains voltage prior to applic...

Page 30: ...inimum separation distance m IEC EN60601 test level V m Compliance level V m 385 380 390 TETRA 400 Pulse modulation b 18 Hz 1 8 0 3 27 27 450 430 470 GMRS 460 FRS 460 FM 5 kHz deviation 1 kHz sine 2 0...

Page 31: ...2 P 80 MHz to 800 MHz d 1 2 P 800 MHz to 2 7 GHz d 2 3 P 0 01 0 12 0 12 0 23 0 1 0 38 0 38 0 73 1 1 2 1 2 2 3 10 3 8 3 8 7 3 100 12 12 23 For transmitters rated at a maximum output power not listed a...

Page 32: ...he following measures Reorient or relocate the receiving antenna Increase the separation between the equipment and receiver Connect the equipment into an outlet on a circuit different from that to whi...

Page 33: ...ZO Corporation All rights reserved 2nd Edition May 7th 2020 Printed in Japan 00N0N165B1 IFU RX1270 6 www eizoglobal com Carl Benz Stra e 3 76761 R lzheim Germany 8 5B 153 Shimokashiwano Hakusan Ishika...

Reviews: