10 Troubleshooting
contra-angle.
Motor is protected by system
or broken.
Check the error warning.
Motor does not
run when the file
is inserted in the
canal.
The measure wire connecting
not properly.
Check the connection.
The lip hook not properly
hooked in the corner of the
patient’s mouth.
Check the connection.
The Auto start function is
OFF
Turn the auto start
function ON
if
necessary.
The motor Can’t
stop.
The Auto stop function is
OFF.
Press main switch to
stop it, setting
Auto stop
function ON if
necessary.
There is a short circuit inside
the motor handpiece or the
motor handpiece cord.
Press “S” button to stop
the motor and c
ontact
your distributor.
/
Motor
spontaneously
starts running in
reverse.
Up to setting torque limit.
Check the torque limit is
enough or not.
Apical
action setting to
Reverse
Change setting if it’s not
expected.
Setting to REV mode.
Change setting if it’s not
expected.
Motor does not
reverse.
R.L mode is set.
Change setting if it’s not
expected.
Torque reverse setting might
be too high.
Change setting if it’s not
expected.
Apical action setting Stop or
OFF.
Change setting if it’s not
expected.
Motor speed
changes
spontaneously.
Apical action setting Slow
Down.
Change setting if it’s not
expected.
Motor alternates
between
forward and
reverse rotation.
Operation mode setting to
REC or ATC.
Change setting if it’s not
expected.
No sound.
Beep volume set to 0.
Set beep volume to 1, 2
or 3.