8 Cleaning
, Disinfection and Sterilization
8.Cleaning, Disinfection and Sterilization
8.1 Foreword
For hygiene and sanitary safety purpose, the components (contra angle, file clip, lip
hook and insulating sleeve) must be cleaned, disinfected and sterilized before each
usage to prevent any contamination. This concerns the first use as well as the
subsequent uses.
Comply with your national guidelines, standards and requirements for cleaning,
disinfection and sterilization.
8.2 General recommendations
8.2.1 The user is responsible for the sterility of the product for the first cycle and
each further usage as well as for the usage of damaged or dirty instruments, where
applicable after sterility.
8.2.2 For your own safety, please wear personal protective equipment (gloves,
safety glasses, etc.).
8.2.3 Use only a disinfecting solution which is approved for its efficacy
(VAH/DGHM-listing, CE marking, and FDA approval) and in accordance with the
DFU of the disinfecting solution manufacturer.
8.2.4 The water quality has to be convenient to the local regulations especially for
the last rinsing step or with a washer-disinfector.
8.2.5 Thoroughly clean and wash the components before autoclaving.
8.2.6 Do not lubricate the motor handpiece.
8.2.7 Do not clean the contra angle with an ultrasonic cleaning device.
8.2.8 Do not use bleach or chloride disinfectant materials.
8.3 Autoclavable Components
Autoclavable Components
Contra Angle
Lip Hook
File clip
Insulating Sleeve
WARNING
●
Only the components above can be autoclaved.
●
Before first use and after each use, sterilize the above components.