Eighteeth E-connect S User Manual Download Page 38

10   Troubleshooting

 

 

10.Troubleshooting 

 

When trouble is found, check the following points before contacting your distributor. 
If none of these are applicable or the trouble is not remedied even after action has 
been taken, the product may have failed. Contact your distributor. 

 

Problem 

Cause 

Solution 

Ref. 
chap 

The power is not 
turned on. 

The battery is flat. 

Charge the battery. 

7.1 

Press the main  switch too 
short time. 

Press the main switch 
more than 0.5 seconds. 

5.1 

The power LED 
on charge base 
does not light. 

Using a wrong adapter. 

Use the original 
adapter. 

4.4 

The adapter is not 
connected. 

Check the connection. 

4.4 

The plug of the adapter is not 
inserted into the outlet. 

Check the connection. 

There is no electricity in the 
outlet. 

Check the connection. 

No charge 
indicator flash 
on handpiece 
screen 

Put the handpiece into the 
charge base in the wrong 
direction. 

Check the direction. 

4.4 

Charge pin of charge base 
unable to rebound. 

Remove debris which 
between move part and 
base of the charge pin. 

Contactors are dirty. 

Cleaning the surface of 
contactors. 

The charge base broken. 

using adapter connect to 
handpiece directly, and 

Contact your 
distributor. 

Handpiece 
screen does not 
appear 

The handpiece broken. 

Check if there is a sound 
of beep or motor, and 

Contact your 
distributor.

 

The motor 
doesn’t rotate. 

M0 mode is stand-alone apex 
locator function. 

Changing to M1-M10. 

6.1 

The contra-angle is clogged 

Clean or replace the 

                                                  Page    38 / 47 

Summary of Contents for E-connect S

Page 1: ...0197 E connect S USER MANUAL...

Page 2: ...6 4 Advanced setting 19 6 5 Parameter logic 21 7 Operation 24 7 1 Charge 24 7 2 Motor operation 25 7 3 Apex operation and not suitable condition 27 8 Cleaning Disinfection and Sterilization 32 8 1 Fo...

Page 3: ...ponents before autoclaving 32 8 2 6 Do not lubricate the motor handpiece 32 8 2 7 Do not clean the contra angle with an ultrasonic cleaning device 32 8 2 8 Do not use bleach or chloride disinfectant m...

Page 4: ...ct S 1 Scope of E connect S 1 1 Parts Identification 1 2 3 5 6 4 7 8 9 1 Charge Base 2 Handpiece 3 Contra Angle 4 Insulating Sleeve 5 File Clip 2 pcs 6 Lip Hook 2 pcs 7 Measuring Wire 8 Adapter 9 Spra...

Page 5: ...gle 1pcs Part No 6036006 Adapter 1pcs Part No 6016007 6016008 Measuring Wire 1pcs Part No 6015011 Lip Hook 2pcs Part No 6072002 File clip 2pcs Part No 6051005 Insulating Sleeve 1pcs Part No 6004027 Sp...

Page 6: ...mber Manufacturer Date of manufacture LOT Lot of manufacture Class II equipment Type B applied part CE marking Direct current WEEE directive marking Keep dry 134 C Can be autoclaved up to a maximum te...

Page 7: ...nto contact with any type of liquids 2 Do not expose the device to direct or indirect heat sources The device must be operated and stored in a safe environment 3 The device requires special precaution...

Page 8: ...it gently to make sure the connection is good The improve insulation of the contra angle during combine apex we recommend using an insulating sleeve You can also use disposable sleeve sold separately...

Page 9: ...t the patient 4 3 Connecting measuring wire If want activity apex measurement function uncap the USB cover on handpiece insert measuring wire Insert lip hook into white slot insert file clip into blac...

Page 10: ...on the screen NOTE Put the handpiece into the charge base in the right direction otherwise the handpiece will not be recharged If only need a base to put the device on dentist element of dental chair...

Page 11: ...then press or wait 5 seconds to confirm Parameter Adjustment Press S till target parameters press or to adjust then press or wait 5 seconds to confirm Preset Program Selection Long press S to entry pr...

Page 12: ...sidual battery power Operation mode Torque of this memory program Unit of torque Newton centimeter Standard working mode The set maximum torque Real time torque Torque display scale The preset speed R...

Page 13: ...ex flash bar Canal measurement start interface Indication number The number has no represent of actual length only for indication Canal length indicator bar Apex flash bar apex stand alone mode or ref...

Page 14: ...locator In the mode the device will work like a stand alone apex locator AP Apex Major apical foramen or Anatomic apical foramen R L Torque reverse less The motor will not reverse rotation no matter...

Page 15: ...special memory for stand alone apex locator function See chapter 7 3 Apex operation and not suitable condition 6 2 Setting parameters WARNING All the parameters must be set according to files make su...

Page 16: ...rque Limit 3 0 N cm The torque setting can be adjusted from 0 5 N cm to 4 0 N cm and R L torque reverse less is also available NOTE The torque of REC and ATC operation mode is difference according to...

Page 17: ...top rotation stop when reach the reference point upward a little bit and will rotate again Off rotating as usual even if reach the reference point Reference point AP 1 3 2 1 During combined length det...

Page 18: ...NOTE ProtaperA RE A GATESA RE A Pro GliderA RE A and Wave oneA RE A is a registered trademark of Dentsply MtwoA RE A Flex MasterA RE A ReciprocA RE A and R PilotA RE A is a registered trademark of VDW...

Page 19: ...M1 or M2 M10 again then press to confirm WARNING Changing the default setting is not recommended only if you are very sure it can be changed otherwise has risk of file broken 6 4 Advanced setting Ver...

Page 20: ...d the left hand can be set Startup memory M1 Press S again the Start memory can be change it means every time turn power on which memory mode will appear first press or to adjust then press to confirm...

Page 21: ...lt memory settings are listed below the setting can be changed as needed Function M1 M2 M3 M4 M5 M6 M7 M8 M9 M10 Operation Mode FWD FWD REC REC ATC ATC REV REV FWD FWD Speed rpm 300 400 350 450 450 30...

Page 22: ...ails are listed below Fwd Rev REC ATC 120 150 200 250 280 300 350 400 450 500 550 600 650 700 750 800 850 900 950 1000 150 200 250 300 350 400 450 500 The torques N cm in different operation mode are...

Page 23: ...0 100 120 150 160 180 200 230 250 260 280 300 320 340 360 370 The same with the front table REV Angle N A The same with the front table The same with the front table NOTE The sum of FWD Angle and REV...

Page 24: ...low power warming will appear on screen and the device will stop work NOTE The remaining amount of battery mark indicates a voltage When a load is applied to the motor handpiece the remaining amount o...

Page 25: ...distributor can change the battery the electronic parts will be damaged if use a wrong battery or install with a wrong way 7 2 Motor operation rpm N cm 3 50 5 4 3 2 1 When using as a stand alone motor...

Page 26: ...tainless steel Gloves and a rubber dam are compulsory during treatment Do not forget to remove the file from the Contra angle after its use 3 1 05 2 AP When using motor combine apex function the measu...

Page 27: ...uggest put the handpiece on the charge base to get better visual angle the measure wire must be connecting with motor by USB socket white slot connects with lip by lip hook and black slot connect with...

Page 28: ...M0 mode check the highlight number must between 01 03 0 3mm 0 8mm on the top OVER 0 3 0 2 0 1 0 0 1 0 4 NOTE We recommend check the connection of apex locator every time before use Touch the lip hook...

Page 29: ...t of the root canal it is impossible to get an accurate measurement It is important to remove the overflow from the opening Build up e g cement Broken crown If the crown is broken a segment of the gin...

Page 30: ...gingival tissue In this case widen the opening at the top of the crown so that the file will not touch the mental prosthesis before taking a measurement Debris Pulp Cutting debris on tooth Pulp inside...

Page 31: ...y image this does not mean inaccurate of apex locator or X ray depending on the angle of the X ray beam the root tip may not be displayed correctly The position of the root tip seems to differ from it...

Page 32: ...2 For your own safety please wear personal protective equipment gloves safety glasses etc 8 2 3 Use only a disinfecting solution which is approved for its efficacy VAH DGHM listing CE marking and FDA...

Page 33: ...le clip Insulating Sleeve before cleaning Refer to Chapter 4 Installing the E connect S of this manual for disassembly instructions WARNING Do not fail to take out the file before cleaning the contra...

Page 34: ...Heat 20min 90 194 F 6 Maintenance and Inspection Inspect components and sort out those with defects Dirty components must be cleaned and disinfected again Only the contra angle needs to be lubricated...

Page 35: ...ait for cooling before touching 9 Storage Keep the components in sterilization packaging in a dry and clean environment WARNING Sterility cannot be guaranteed if packaging is open damaged or wet Check...

Page 36: ...ng Wire Handpiece Base Wipe all the surfaces with a cloth lightly moistened with Ethanol for Disinfection Ethanol 70 to 80vol at least 2 min repeat for 5 times NOTE Do not use anything except Ethanol...

Page 37: ...ress the Main switch to restart motor Overheat See user manual The temperature of motor is higher than expectation turn the power off and waiting more than 5 minutes to let it cold down HWFault See us...

Page 38: ...The plug of the adapter is not inserted into the outlet Check the connection There is no electricity in the outlet Check the connection No charge indicator flash on handpiece screen Put the handpiece...

Page 39: ...he motor and contact your distributor Motor spontaneously starts running in reverse Up to setting torque limit Check the torque limit is enough or not 6 2 Apical action setting to Reverse Change setti...

Page 40: ...surement is unstable Complex root canal environment Check situation of root canals 7 3 Cannot make a measurement Measure wire lip hook or file clip connecting not properly Check the connection 7 2 7 3...

Page 41: ...mAh 10 Charger power supply AC 100 240 V 10 Charger power output 5V 1A Frequency 50 60Hz 10 Charger nominal power input 5 5VA Torque range 0 5N cm 4N cm Speed range 120 1000 rpm Type of protection aga...

Page 42: ...tic establishments and those directly connected to the public low voltage power supply network that supplies buildings used for domestic purposes Harmonic emissions IEC61000 3 2 Class A Voltage fluctu...

Page 43: ...Mains power quality should be that of a typical commercial or hospital environment If the user of devices require continued operation during power mains interruptions it is recommended that devices be...

Page 44: ...C 61000 4 3 3V m Proximity fields from RF wireless communication equipment Complies IEC 61000 4 3 Recommended minimum separation distances Nowadays many RF wireless equipments have being used in vario...

Page 45: ...802 11 a n Pulse modulation 217Hz 0 2 0 3 9 5500 5785 WARNING 1 Use of accessories and cables other than those specified or provided by the manufacturer of E connect S could result in increased elect...

Page 46: ...mponents and printed circuit boards are disposed as electrical scrap The lithium batteries are disposed as special refuse Please deal with them according to the local environmental protection laws and...

Page 47: ...hou City 213000 Jiangsu China Tel 86 0519 85962691 Fax 86 0519 85962691 Email ivy sifary com Web www sifary com Llins Service Consulting GmbH Tel 49 175 4870819 Add Obere Seegasse 34 2 69124 Heidelber...

Reviews: