6 Setting
(without insulating sleeve) or file touch the
patient’s lip or operator’s fingers (without insulating
glove), take care to avoid this, the file rotated by
motor has risk of injure someone.
Aut o St op
OFF
When the endo file out of root canal, the motor will
not auto stop with default setting, Press
<
or
>
to select auto stop “ON” if needed.
Apical A ct ion
Reverse
Because of
integrated apex locator, when the file
reaches the reference point, the motor will
response according to setting, it can be Reverse,
SlowDown, Stop and Off.
Press
<
or
>
to change.
Reverse:
rotation direction changing till the file
upward a little bit by operator, rotation direction will
change back again.
SlowDown:
rotation slowdown when approach the
reference point, will reverse if reach.
Stop:
rotation stop when reach the reference
point, upward a little bit and will rotate again.
Off:
rotating as usual even if reach the reference
point.
Reference point
AP
1
3
2
1
During combined length determination, normally
apical reverse must active before reaching major
apical foramen,
Press
<
or
>
to set apical
reverse position by change the flash bar (
A
○
1
E
A
), the
motor will reverse while reaching the flash bar
every time.
FWD A ngle
120
°
Activating in REC and ATC operation mode.
forward angle (Clockwise rotation angle) can be
adjusted by operator from 30° to 370°,
Press
<
or
>
to change.
REV A ngle
150
°
Activating in REC and ATC operation mode.
reverse angle (Counter Clockwise rotation angle)
can be adjusted by operator from 30° to 370°,
Press
<
or
>
to change.
NOTE
The sum of FWD Angle and REV Angle must be
greater than 120°, the motor system has closed
the angle not needed. For example: if you set
FWD Angle 30°, the REV Angle must be setting
greater than 90°.
Page 17 / 47