6001
Write-protect switch on SD card activated. Deactivate
this switch on the SD card
6002
SD card not inserted
6003
Write protection switch SD card activated
6010-6012
Data set could not be saved to SD card.
Repeat weighing, if the error occurs again, check SD
card.
6013
Storage not possible, time was not set.
Set a valid time.
6055
Transmission of the measurement data to the USB port
canceled by pressing the Off button on the DRC 433.
6100-6238
Measurement data on SD card incorrect, unusable.
The measurement data were manipulated with a PC.
The measurement data are fragmentary due to a
defective SD card.
No data connection to program Trans
Use a COM port between 1-9.
Wrong BAUD rate set.
DRC 433 has no connection to the scale.
Measurement data cannot be read out.
Commands on the USB port are only accepted:
1.in the pause between 2 consecutive records, or
2. if the USB port data transfer is disabled. See
instructions par. data protocol 28 bytes EHP scales, or
3. when the scale is switched off.
No weighing data at the VCP
Data transfer on the USB port has been deactivated.
See also instructions Para. data protocol 28 byte EHP
scales
Reception display flashes, but no
weight display
Scale number set incorrectly. Channel is set correctly.
Receive indicator does not flash
rhythmically at 1 second intervals.
The reception situation is poor. Select another radio
channel, see channel table.
Tare is not evaluated in the display,
only a line is visible in the tare line.
Old scales firmware. Contact the EHP service.
Incorrect data protocol EHP scales. Contact EHP Service.