The DRC 433 works with error codes. Error codes that you may be able to correct yourself are listed
below. If you get an extended system error displayed that is not listed here, please contact EHP Service.
No function
Check the batteries/rechargeable batteries.
Press and hold the power button for a longer time.
The display shows "Scale no.
incorrect".
The scale number was set incorrectly in the setup or
another EHP scale is transmitting on the same radio
frequency
The display shows "Error no. xxxx".
(xxxx corresponds to an error number)
Press the OK key to acknowledge the error. The DCR
433 will now restart automatically.
The meaning of the respective error number is listed in
the following lines.
4
Modem error - if the error is not corrected after a
restart, contact EHP Service.
5
USB installation faulty - reinstall driver, check USB
connection on PC
6
SD card error - Format the SD card on your PC. If the
error persists, use another SD card. Do not use SDHC
cards.
1162
An incorrect channel has been selected for High Range
DRC 433. Select a channel between 1-28.
1165
An incorrect channel has been selected for Low Range
DRC 433. Select a channel between 30-39.
1166
An incorrect channel has been selected for Low Range
DRC 433. Select a channel between 30-39.
For devices with High Range radio, the modem does not
initiate. Remove the batteries from the device for 3
minutes and check again.
1170-1181
Low Range Modem defective/ faulty. Contact EHP
Service.
6000
SD card error - Format the SD card on your PC. If the
error persists, use another SD card. Do not use SDHC
cards.