![EEC ESD-140 Operation Manual Download Page 37](http://html1.mh-extra.com/html/eec/esd-140/esd-140_operation-manual_2366405037.webp)
33
Bond Tester. Braces ({ }) enclose each parameter for a command string. Triangle
brackets (< >) indicate that you must substitute a value for the enclosed parameter.
The Pipe ( | ) is used to separate different parameter options for a command. The
command and the parameter data must be separated with a space. All commands
that end with a question mark (?) are query commands.
Test Execution Commands
The following commands are used to control actual output voltage and current from
the instrument. Please observe all safety precautions.
COMMAND
DESCRIPTION
TEST
Execute a Test
RESET
Abort a Test in Process or Reset
Failures
AOFF
Auto Set Offset
TEST
Starts the test sequence at the selected step loaded into memory (RAM).
RESET
Stop or abort a test. Also used to reset a latched failure condition.
AOFF
Set the offset for the ground bond test. The cables and any test fixture should be
connected before executing the command. This command will perform an actual test
and all safety precautions should be observed when using this command.
File Editing Commands
The following commands are used to create or modify Test Setup Files.
COMMAND
DESCRIPTION
VALUE
MSS <
memory number step
number
>
Memory Step
Select
memory number
= 1-10
step number
1-3
SALL<
step,p1,p2,p3…>
Add all
Parameters of a
Test
MSS<memory number step number>
Load a file by memory and step number. The first digit is the memory number and the
second digit is the step number. Example: MSS 11 = memory 1 step 1
SALL <step, p1, p2, p3… >
This command edits all parameters in a memory and step. Parameters will be edited
at the step location that has been selected. Example: SALL<
11,10.00,100,0…>
The parameter <
step>
indicated the memory and step number. The
parameters <
p1,p2>
etc. indicate the individual settings for each test. All parameters