31
Hipot Start parameter
The Hipot Start parameter, found in the System Menu screen, controls whether the
Hipot test will run sequentially after the Ground Bond test or simultaneously with the
Ground Bond test. Please refer to section
4.1 Setup System
for instructions on how
to set this parameter.
The Hipot Start parameter is only used when connecting the Ground Bond Tester with
an eec HIPOT Tester. The parameter determines if the Hipot will run after the Ground
Bond Tester has completed its test, G
W, or if the Hipot will run at the same time as
the Ground Bond Tester, G + W.
The Hipot Start parameter directly controls the Start output relay. Setting the Hipot
start parameter to G
W will momentarily close the start output relay at the end of the
Ground Bond test. Setting the Hipot start parameter to G + W will momentarily close
the start output relay at the start of the Ground Bond test