![Edwards T-STATION 85 Instruction Manual Download Page 23](http://html.mh-extra.com/html/edwards/t-station-85/t-station-85_instruction-manual_2366171023.webp)
Figure 9 Turbo setpoint number entry mode
Turbo setpoint screen key actions
Table 8 Turbo setpoint screen key actions
Keys
Short press
Long press
START
Go to turbo screen
-
ENTER
Enter edit mode, or confirm
selection
Enter number entry mode
NEXT
Next TSP option, next number or go
to gauge calibration screen
Scroll through TSP options in edit
mode, scroll through numbers in
number entry or scroll through
menu screens
Calibrate gauge screen
When the calibrate gauge screen is selected, the gauge LED and the CAL LED are lit together.
The numeric display is blank. The calibrate gauge menu is not available for gauges which do not
support calibration.
When the ENTER key is pressed the action depends on gauge type:
WRG or APGX: The calibration command is sent to the gauge and the display shows "CALd" for
3 seconds.
ASG: The calibration functions as a zero offset adjustment. The pressure currently displayed is
saved as the zero offset and is subtracted from all future readings. The display shows "CALd"
for 3 seconds to confirm the action. The offset adjustment can be cancelled by pressing the
ENTER key again. The display shows "OFF" for 3 seconds to confirm that the offset adjustment
has been removed.
D395-94-880A - Operation