D0300040001EN - Parameters
Page 33
Sensor switch-off value (
)
This parameter will only appear when the sensor switch-off type has been set to
or
(see
Sensor switch-off type (S-off)
).
Through the parameter
you can define a switch-off value. When the pressure in the
respective measurement channel exceeds the switch-off value, the gauge head is switched
off.
General parameters (
)
With the aid of these parameters you can generally configure the instrument. The parameters
apply to all measurement channels.
Keep the pushbutton (PARA) depressed for approximately 2 seconds.
The instrument is now in the parameter mode.
Use the arrow pushbuttons so as to access the parameter group
.
Press the pushbutton (PARA) to select the desired parameter.
The name and the value of the parameter are displayed.
Use the arrow pushbuttons to change the parameter setting.
Repeat the steps to change further parameters of the parameter group.
Unit of measurement (
)
Unit of measurement for pressure values. This unit of measurement affects the displayed
pressure values, threshold values etc.
Manual. The gauge head can be switched off by pressing the pushbutton
(DOWN) (see
Externally through an optocoupler (PLC compatible low level < 7 V d.c.; 0 A at
24 V d.c.).
Self-monitoring. With the aid of the next parameter
, a switch-off value
can be defined. When the pressure at the gauge head exceeds the switch-off
value then the gauge head is switched off.
Through measurement channel PRG 1. With the aid of the next parameter
, a switch-off value can be defined. When the pressure in the
measurement channel PRG 1 exceeds the switch-off value then the sensor is
switched off.
Through measurement channel PRG 2. With the aid of the next parameter
, a switch-off value can be defined. When the pressure in the
measurement channel PRG 2 exceeds the switch-off value then the sensor is
switched off.
Summary of Contents for PGC201
Page 10: ...This page has been intentionally left blank D0300040001EN Contents Page vi ...
Page 14: ...D0300040001EN Safety Page 4 This page has been intentionally left blank ...
Page 30: ...D0300040001EN Installation Page 20 This page has been intentionally left blank ...
Page 38: ...D0300040001EN Operation Page 28 This page has been intentionally left blank ...
Page 46: ...D0300040001EN Parameters Page 36 This page has been intentionally left blank ...
Page 58: ...D0300040001EN Computer interface Page 48 This page has been intentionally left blank ...
Page 62: ...D0300040001EN Storing and waste disposal Page 52 This page has been intentionally left blank ...
Page 63: ...This page has been intentionally left blank ...
Page 64: ...This page has been intentionally left blank ...