Technical data
B800-00-880 Issue D
2.9
Controller connector socket
The nEXT pump has a 3-way controller connector socket in the side of the controller. When the pump is shipped, this
connector is concealed by a black protective cover. If the connector will be used, the cover should be removed by
levering with a small screwdriver. The mating plug for this connector is supplied with the pump.
The connector is intended to drive a vent valve or fan connected to the three pins. The connector mating plug is
shown in
Figure 9
, with the polarity of the pins marked when the vent valve/fan is energised.
The 24 V d.c. output on the controller connector is protected against overload or short circuit to 0 V d.c./24 V d.c.
Rtn/chassis. The output is guaranteed to supply the current given in
Table 12
, but for higher currents the output
voltage will be reduced and may turn off completely to limit current.
Figure 9 - Controller connector showing pin numbers
Table 12 - Controller technical data
Description
Data
Connector plug
Phoenix part number FK-MC1881338
Voltage output
24 V d.c. -25%, +10% (18 V d.c. to 26.4 V d.c.)
Current output
500 mA
Pin number
Signal
Polarity
1
24 V d.c. Out
+
2
Chassis
3
24 V d.c. Rtn
-