INSTALLATION
A533-23-880 Issue C
3.5
Locate the iF system
Use the following procedure to locate the iF system in its operating position. The iF system must be located on a
firm, level surface.
1. Use suitable lifting equipment attached to the lifting bolts to move the iF system close to its final operating
position.
2. Wheel the iF system on its castors to move it into its operating position.
3. Adjust the levelling feet to make sure that the iF system is level and is not supported by the castors.
3.6
Check the pump oil level(s)
CAUTION
Ensure that the oil-levels in the FDP pump and the HMB pump are correct. If a pump oil-level is incorrect, pump
performance may be affected and the pump may be damaged.
3.6.1
iF1800 Systems
The iF1800 systems are supplied with both the FDP and HMB pumps filled with oil. However, we recommend that you
check the FDP and HMB oil-levels as described below.
1. Check the FDP pump oil-level.
2. Refer to Figure 1. Check that the oil-level is between the MAX and MIN marks on the bezel of the HMB oil-level
sight-glass (5):
If the oil-level is above the MAX mark, drain excess oil from the pump as described in Section 6.10 until the
oil-level is correct.
If the oil-level is below the MIN mark, fill the pump with oil until the oil-level is correct, as described in
Section 6.10.
WARNING
You must use suitable lifting equipment to move the iF system. It is too heavy to lift by hand.
WARNING
Ensure that the maximum angle between paired slings used to lift the iF system is 45
o
, and that all
four lifting bolts are used.
WARNING
Ensure that the maximum angle between paired slings used to lift the iF system is 45
o
, and that all
four lifting bolts are used.