Chapter 2 — The Quad Tc/mV Amp
13
Manual Calibration
If known temperature sources are available then very accurate
calibration can be achieved using the raw mV signal setting. The signal
can be calibrated over small ranges with a linear function using the
Chart software Units Conversion feature (see the
Chart Software
Manual
). Over wider ranges the Multipoint Calibration extension
software plug-in can be used, see
http://www.edaq.com/chart_extns_list.html
Tc
Type
Temperature Range
Limits of Error
°C
°F
Standard
Special
T
–200 to –66
–328 to –87
±1.5%
—
–66 to 0
–87 to +32
±1°C, ±2°F
—
0 – 133
32 – 270
±1°C, ±2°F
±0.5°C, ±1°F
133 – 350
270 – 662
±0.75%
±0.4%
J
0 – 293
32 – 559
±2.2°C, ±4°F ±1.1°C, ±2°F
293 – 750
559 – 1382
±0.75
0.4%
E
–200 to –100
–328 to –148
±1%
—
–100 to 0
–148 to +32
±1.1°C, ±3°C
—
0 – 340
32 – 644
±1.7°C, ±3°F
±1°C, ±2°F
340 – 900
644 – 1652
±0.5%
±0.4%
K
–200 to –110
–328 to –166
±2%
—
–110 to 0
–166 to +32
±2.2°C, ±4°F
—
0 – 293
32 – 559
±2.2°C, ±4°F ±1.1°C, ±2°F
293 – 1250
559 – 2282
±0.75%
±0.4%
N
0 – 293
32 – 559
±2.2°C, ±4°F ±1.1°C, ±2°F
0 – 1250
559 – 2282
±0.75%
±0.4%
R, S
0 – 600
32 – 1112
±1.5°C, ±3°C ±0.6°C, ±1°F
600 – 1450
1112 – 2642
±0.25%
±0.1%
B
870 – 1700
1598 – 3092
±0.5%
—
Table 2–2
Limits of Error for
Thermocouples,
Reference Junction
0 °C (32 °F)