SE-3&SE-300 Series Electrocardiograph User Manual Settings
- 54 -
Select
Age Group
, the
Age Group
textbox appears in the
Patient
Information
window.
PatInfo Refreshed
Choose from:
On
or
Off
,
the default value is
On
Select
On
, the patient information will be refreshed after the ECG report
is printed out and all the leads are off.
H/W Unit
Choose from:
cm/kg
or
inch/lb
,
the default value is
cm/kg
BP Unit
Choose from:
mmHg
or
kPa
,
the default value is
mmHg
8.5 Transmission Setup
NOTE
:
1.
To transmit ECG data to the PC in DAT format, the Smart ECG Viewer software
produced by the manufacturer must be installed in the PC. You should log into the
Smart ECG Viewer software before transmission.
2.
To transmit ECG data to the PC in DICOM/SCP/FDA-XML/PDF format, the FTP
receiving software must be installed in the PC. You should log into the FTP receiving
software before transmission.
Item
Description
Device No.
Enter Device No., it should be within 7 ASCII characters.
Auto Transmission
Choose from:
On
or
Off
,
the default value is
Off
Select
On
, ECG data will be transmitted automatically after an ECG
report is printed out in the auto or rhythm mode; in the off mode, 10s
ECG data sampled before pressing the
PRINT/STOP
key can be
saved and transmitted, but cannot be printed.