SE-3&SE-300 Series Electrocardiograph User Manual Operation Preparations
- 27 -
Table 4
–1 Electrodes and Their identifiers and color codes
European
American
Electrodes
Identifier
Color code
Identifier
Color code
Right arm
R
Red
RA
White
Left arm
L
Yellow
LA
Black
Right leg
N or RF
Black
RL
Green
Left leg
F
Green
LL
Red
Chest 1
C1
White/red
V1
Brown/red
Chest 2
C2
White/yellow
V2
Brown/yellow
Chest 3
C3
White/green
V3
Brown/green
Chest 4
C4
White/brown
V4
Brown/blue
Chest 5
C5
White/black
V5
Brown/orange
Chest 6
C6
White/violet
V6
Brown/violet
As the following figure shows, the positions of chest electrodes on the body surface are
C1: Fourth intercostal space at the right border of the sternum
C2: Fourth intercostal space at the left border of the sternum
C3: Fifth rib between C2 and C4
C4: Fifth intercostal space on the left midclavicular line
C5: Left anterior axillary line at the horizontal level of C4
C6: Left midaxillary line at the horizontal level of C4
C6
C1
C2
C5
C4
C3