SE-3&SE-300 Series Electrocardiograph User Manual Managing ECG Records
- 41 -
3) There should be no shield in or around the room where the wireless AP is used, or
else the wireless transmission may fail.
Auto Transmission:
1.
Press the
MENU
key to open the System Setup Screen.
2.
Select the
Transmission
button, and then press the
MENU
key to enter the sub-window.
3.
Set
Auto Transmission
to
On
and set
Transmission mode
to
Ethernet.
4.
Set the
Server IP
to the IP of Smart ECG Viewer.
5.
Set the first three numbers of the
Local IP
to the first three numbers of the IP of Smart
ECG Viewer. The last number of the
Local IP
item can be set at random, but it can’t be the
same as the last number of the IP of Smart ECG Viewer.
For example, check the IP on the
System Setting
interface of the Smart ECG Viewer
software.
Set the
Server IP
to the IP of Smart ECG Viewer. Set the first three numbers of the
Local
IP
to the first three numbers of the IP of Smart ECG Viewer. The last number of the
Local
IP
can be set at random, but it can’t be the same as the last number of the local IP of Smart
ECG Viewer.
6.
In the
AUTO, RHYT, OFF-AUTO
or
OFF-RHYT
mode, ECG data can be transmitted
through the net automatically after an ECG report is printed out.
Manual Transmission:
1.
Press the
MENU
key to open the System Setup Screen.
2.
Select the
Transmission
button, and then press the
MENU
key to enter the sub-window.
3.
Set
Auto Transmission
to
Off
and set
Transmission mode
to
Ethernet.
4.
For IP address setting, refer to step 4 and 5 for auto transmission.
5.
Other steps for transmission operation, refer to step 4 for manual transmission in section
7.1.1 “Transmitting ECG Record through the Serial Port”.
NOTE:
The transmission process is long, and please be patient to wait.