VE-1010 Veterinary PC ECG User Manual EMC Information
- 69 -
Guidance and manufacture’s declaration - electromagnetic immunity - for EQUIPMENT
and SYSTEMS that are not LIFE-SUPPORTING
Guidance and manufacture’s declaration - electromagnetic immunity
VE-1010 VET PC ECG is intended for use in the electromagnetic environment specified below.
The customer or the user of VE-1010 VET PC ECG
should assure that it is used in such an
environment.
Immunity
test
IEC 60601 test level
Compliance
level
Electromagnetic environment -
guidance
Conducted RF
IEC61000-4-6
Radiated RF
IEC61000-4-3
3 V
rms
150 kHz to 80 MHz
3 V/m
80 MHz to 2.5 GHz
3V
rms
3 V/m
Portable and mobile RF
communications equipment should be
used no closer to any part of VE-1010
VET PC ECG, including cables, than
the recommended separation distance
calculated from the equation applicable
to the frequency of the transmitter.
Recommended separation distance
P
d
2
.
1
P
d
2
.
1
80 MHz to 800 MHz
P
d
3
.
2
800 MHz to 2.5 GHz
Where
P
is the maximum output power
rating of the transmitter in watts (W)
according to the transmitter
manufacturer and
d
is the
recommended separation distance in
metres (m).
Field strengths from fixed RF
transmitters, as determined by an
electromagnetic site survey,
a
should be
less than the compliance level in each
frequency range.
b
Interference may occur in the vicinity
of equipment marked with the
following symbol:
NOTE 1 At 80 MHz and 800 MHz, the higher frequency range applies.
NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation
is affected by absorption and reflection from structures, objects and people.
a
Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless)
telephones and land mobile radios, amateur radio, AM and FM radio broadcast and TV