VE-1010 Veterinary PC ECG User Manual Hint Information
- 58 -
Chapter 9 Hint Information
Hint information and the corresponding causes provided by the system are listed as follows.
Table 9-1 Hint Information and Causes
Hint Information
Causes
Lead off: X
Electrodes fall off the pet or the patient cable
falls off the ECG sampling box.
It is pre-sampling now, please click on
“Start” to begin recording.
During the pre-sampling course
Resting ECG is sampling now!
During the sampling course of Resting ECG
Can't detect the Sentinel, enter DEMO?
The sentinel is not connected.
Sentinel is not compatible, enter DEMO?
The sentinel is set wrongly.
Hint: Please make sure the USB cable has
been connected. If possible, please
re-connect it!
The USB cable is disconnected or the
communication between the ECG sampling box
and the serial port is interrupted.
1. Reconnect the ECG sampling box to the PC.
2. Click on the
Device
tab on the
System
Setting
interface of the VET PC ECG system,
and check whether the sampling device is set
correctly.
Communication error! Please check the
USB cable!
The USB cable falls off the PC during the
sampling process.
Overload
The direct current offset voltage on an electrode
is too high.
Sorry, Can not Connect to the Database!
MSDE 2000 or SQL Server 2005 Express is not
started up.
Fail to create database!
The system fails to create database.