Table 2.2 (continued)
Trouble
Probable
cause
Solution
2 Low
battery
indication is
displayed
when the
detector is
switched on
after the
batteries
have been
replaced
1 Poor contact
between the
batteries and
the battery
compartment
clamps
2 One of the
batteries is out
of order
1 Clean out the
contacts on the
clamps and the
batteries
2 Replace the
batteries
2.3.2.2 If troubles listed in Table 2.2 fail to be
eliminated, or more complicated problems occur, the
detector should be sent for repair to appropriate repair
services or to the manufacturer (see Repair section).