26
reset to ‘off’ if 1.16 is set to ‘on’.
1.16
0…50 μS/cm conductivity range
Specify if the controller will read electrical conductivity of
permeate in the range of 0…50 μS/cm (on/off). This setting will reset
to ‘off’ if 1.15 is set to ‘on’.
1.17
Permeate conductivity Fault threshold
Specify maximum acceptable permeate conductivity. Conductivity
reading above this value will initiate Fault
mode (‘High permeate
TDS’). If set to zero,
fault threshold will not be used.
1.18
Permeate conductivity Fault delay
Specify the length of time before the controller goes into Fault
mode when high permeate conductivity is being read.
1.19
Temperature probe
Select whether temperature probe is used (on/off).
1.20
Set permeate temperature
Specify temperature of permeate for correct conductivity
measurement. Only active if 1
.19 is ‘off’.
1.20
Divide temperature by 10
Temperature
value ÷10
will be shown
. Only active if 1.19 is set to ‘on’
.
1.21
New settings and calibration passcode
Verify passcode.
2.
CALIBRATION
Hold
STOP for 8 seconds to launch menu prompt. Push
STOP
to skip Settings submenu and push
START to enter Calibration
submenu. Enter passcode in the prompt. Factory passcode is ‘0000’.
2.1
First point value
First calibration point can be done at zero electrical conductivity (dry
conductivity meter). In order to use zero first point conductivity, remove
the conductivity meter from its cell, wipe with clean cloth and keep dry
for a few minutes. When conductivity reading on the display stabilizes,
put zeroes in the bottom row, and go to the next step.
If using a weakly conducting solution to set the first point, rinse the
meter with deionized water and wipe dry. Dip clean conductivity meter