Page 27
11.4. Soot...Oil trace (option)
The
sub-menu
"Soot...Oil trace"
enables the input of measured results
for boiler temperature, soot dots and oil
trace. Select the line
„Boiler temp.“
and press
<OK>
to activate the input.
The input can be made using the
numerical keys. Press
<OK>
to store
the value in the data record of the
measurement.
The soot dot measurement is to be performed with the optional
heated pistol grip probe which heating function prevents the filter
paper to become wet because of the humidity issued by the
combustion condensate. The filter paper slot is hereby heated up to
approx. 70 °C. Switch hereto the probe heating of the pistol grip
probe while selecting
„Adjustments / Internal / Probe heating /
<F1>“
.
Proceed as follows:
-Switch on the probe heating while selecting
„Adjustments / Probe
heating / <F1>“
.
-Insert a filter paper in the paper slot.
-Select the line
„1st. Soot meas.“
.
-Press
<OK>
to start the measurement. The display shows the
volume to be sucked and the pump starts sampling.
If the soot dot analysis are made with a manual pump the sucking
procedure can be interrupted while pressing
<F4>
(result value can
immediately be entered).
Once 1,63 litres has been sucked in, the instrument will instruct to
input the opacity degree. Proceed as follows:
-Release the filter paper from the probe slot.
For this option a probe
type SB is required!
Soot..Oil trace
Mean value: -.-
Boiler temp.: 66°C
1st Soot meas.: -.-
2nd Soot meas.: -.-
3rd Soot meas.: -.-
Oil trace : ----