WLAN, Display values, Pressure. Further you can switch with
<F1>
from any menu back to the measured value display.
The position of the measured and calculated values (gas analysis)
on the display pages is free selectable (choose
„Display values“
for
<F4>
Hotkey). For alteration of the existing succession respectively
personal listing, proceed as follows:
-Press
<F4>
to activate the function.
-select the line with the cursor keys
<up/down>
,
-select the measured or calculated value with the cursor
keys
<right/left>
,
-repeat this procedure until all modifications are completed.
-Press
<F4>
to deactivate the function.
Instruments with magnetic valve and fresh air connection
Instruments that are equipped with magnetic valve technology and
fresh air connection are able to be purged or calibrated with fresh air
without taken out the probe out of the flue gas duct:
For fresh air purging, press the
<Enter>
key from the measured
value display, select the menu item
"Fresh air purge"
and press the
<Enter>
key again. The fresh air purging can be ended with
<F3>
in
the measured value display.
For fresh air calibration, press the
<Enter>
key from the measured
value display, select the menu item
"Re-Calibration"
and press the
<Enter>
key again. After the following query has been answered
with
"YES"
(
<F1>
), the 1-minute calibration phase starts.
Core stream search
Position the sampling probe in the exhaust channel so that the
thermocouple is fully surrounded with the gas.