-Soot measurement calibration
-Connect suitable reference principle for volume determination
-Insert filter paper
-Confirm with
<OK>
-Start soot measurement with
<OK>
-After 1,63 l volume has been sucked, press again
<OK>
(operation time of the pump will be memorised)
-Press
<ESC>
to return to the main menu
Unit
(adjustment with
cursor keys
):
-Calculation of gas concentrations in:
-ppm
= volume concentration (parts per million)
-mg/m
3
= mass concentration per volume unit
-mg/kWh (undiluted) = mass concentration per power unit
-mg/MJ (undiluted)
= mass concentration per power unit
-ppm (undiluted)
= volume concentration (parts per million)
-mg/m
3
(undiluted) = mass concentration per volume unit
Undiluted
:
-Conversion of the gas concentration on selected reference oxygen:
-mg/kWh and mg/MJ are always calculated on 0% O2 basis
-Conversion formula:
E
ref
= E
meas
*
21 - O
2ref
21 - O
2meas