Page 27
14. Data Processing (continued)
Proceed as follows:
-Connect ecom-EN2-F and PC via USB cable.
-Select
“Load data“
and confirm with
<Enter>
.
-Answer the displayed question with
YES
(
<F1>
).
-Decide if the data recorded can be cancelled
(
<F1>
for
YES
/
<F4>
for
NO
).
-Start the data transfer on your PC.
Send data:
With this function the data records completed with measurement
values can be transferred to the PC programme (procedure similar to
chapter
„Load data“
).
15. Delta-T Measurement (Option)
With the ecom-EN2-F a difference temperature measurement is pos-
sible. For measurements at pipings (e.g water-in and water-out of
heating systems), special temperature sensors are needed, available
from your authorised rbr agency. Out of the main menu point
"Diag-
nostics"
select the sub-menu
"dT measurement"
and confirm with
<Enter>
.
The instrument indicates the tempera-
ture T1 (sensor at connection „Gas
temperature“), the temperature T2 (sen-
sor at connection „Air temperature“) and
the difference between both tempera-
tures (T1 - T2). Press
<Memory>
to
store the result in the intermediate
memory. A printout can be started with
<Print>
.
Measurement stored in
intermediate memory
dT measurement
T1
70.4
°C
T2
56.3
°C
dT
14.1
°C