Page 11
5. Power Supply
The EN2-F is recommended to be powered by AC power when
available. The analyzer can also be powered using the internal bat-
tery for a while (6 V; 3.8 Ah). The battery should be recharged when
the analyzer starts beeping or an alert message appears. The bat-
tery life can be checked by looking at the voltage indication on the
display in the
Control
menu. The low battery warning is activated
when the value
“Batt” falls below 5.7 V. When the analyser is con-
nected to AC power, the voltage will read as dashes “—.—“ Unplug
the power cord to read the battery voltage.
When the battery voltage falls below 5.4 V, the instrument must
be connected to AC power to retain power.
6. Data Logging onto Memory Card
The multi-media card enables the storage of data records on a re-
movable SD card. The data fields on the csv file are listed at the end
of this manual. The following conditions must be fulfilled for using a
multi-media card:
- minimal card volume 32 MB - max. 2 GB
- card formatted on 16 bit FAT
Insert Memory Card
Insert the multi-media card as shown.
Take care that the card inserts smoothly
and that you do not jam it in.
Used batteries can be brought to recycling stations or
public waste disposal companies!
Never pull out cards during data recording
as this may cause a loss of data and/or
damage to the memory card itself.